The EHE010 is intended for use in the electromagnetic environment specified below. 
The customer or the user of the EHE010 should assure that it is used in such an environment.

RF emissions
CISPR 11

RF emission
CISPR 11

Harmonic emissions
IEC 61000-3-2

Voltage fluctuations/ 
flicker emissions
IEC 61000-3-3

Group 1

The EHE010 use RF energy only for its internal 
function. Therefore, its RF emissions are very low 
andare not likely to cause any interference in 
nearby electronic equipment.

Class B

The EHE010 is suitable for use in all 
establishments, including domestic 
establishments and those directly connected to 
the public low-voltage power supply network 
that supplies buildings used for domestic 
purposes.

Not applicable
(internal battery 
powered)

Not applicable
(internal battery 
powered)

EMISSION TEST

Guidance and manufacture’s declaration – electromagnetic emission

COMPLIANCE

ELECTROMAGNETIC ENVIRONMENT - 
GUIDANCE

Electrical fast 
transient/burst
IEC 61000-4-4

±2 kV for power 
supply lines
±1 kV for input/
output  lines

Not applicable
(internal battery 
powered)

Not applicable
(internal battery 
powered)

Not applicable
(internal battery 
powered)

Mains power quality should be that of a
typical commercial or hospital 
environment.

Mains power quality should be that of a
typical commercial or hospital 
environment.

Surge
IEC 61000-4-5

± 1 kV line(s) 
to line(s)
± 2 kV line(s) 
to earth

Voltage dips, short 
interruptions and 
voltage variations 
on power supply 
input lines
IEC 61000-4-11

<5% UT (>95% dip 
in UT) for 0.5 cycle

40% UT (60% dip 
in UT) for 5 cycles

70% UT (30% dip 
in UT) for 25 cycles

<5% UT (>95% dip 
in UT) for 5 sec

Mains power quality should be that of a 
typical commercial or hospital 
environment. If the user of the EHE010 
requires continued operation during 
power mains interruptions, it is 
recommended that the EHE010 be 
powered from an uninterruptible power 
supply or a battery.

Power frequency
 (50Hz/60Hz) 
magnetic field 
IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields 
should be at levels characteristic of a 
typical location in a typical commercial 
or hospital environment.

NOTE:  UT is the a.c. mains voltage prior to application of the test level.

Electrostatic 
discharge (ESD)
IEC 61000-4-2

±6 kV contact
±8 kV air

±6 kV contact
±8 kV air

Floors should be wood, concrete or 
ceramic tile. If floor is covered with 
synthetic material, the relative humidity 
should be at least 30%.

IMMUNITY
TEST

IEC 60601 
TEST LEVEL

COMPLIANCE
LEVEL

ELECTROMAGNETIC 
ENVIRONMENT - GUIDANCE

The EHE010 is intended for use in the electromagnetic environment specified below. 
The customer or the user of the EHE010 should assure that it is used in such an environment.

Guidance and manufacturer’s declaration – electromagnetic immunity

14

15

Summary of Contents for EHE010

Page 1: ...imulator User Manual Questions or Comments Please call toll free 1 855 822 6999 M F 9 a m 5 p m CST E mail service healthcare manager com Easy Healthcare Corporation 360 Shore Dr Burr Ridge IL USA 605...

Page 2: ...Procedure Mode Reference Chart 5 Attaching Electrode Pads to Body 6 TENS Electrode Placement Suggestions 7 Best Practices 7 FAQ 8 Specifications 9 Safety Warning 10 Electromagnetic Compatibility and F...

Page 3: ...remaining through clear visual representations It has a smooth soft touch keypad and lock unlock feature These additional features enhance your user experience so that you can use your tens device an...

Page 4: ...witch Knob to adjust the intensity of either independent channel A independent channel B or both Press TIMER to adjust the time Choose from 20 40 or 60 minutes let off and press again to get to next t...

Page 5: ...to bring comfort and relief Experiment with the placement of the pads to find the position that brings the most relief to you Although place pads around the painful area and not directly on it FOREAR...

Page 6: ...s to the proper intensity shown in the round intensity window for either side FAQ Why am I not experiencing any results While you might feel refreshed after your first treatment it takes time for long...

Page 7: ...est may cause rhythm disturbances to the heart which could be lethal 2 over painful areas Please consult with your physician before using this device if you have painful areas 3 over open wounds or ra...

Page 8: ...operation of the unit 3 Caution This unit has been thoroughly tested and inspected to assure proper performance and operation 4 Caution This machine should not be used adjacent to or stacked with oth...

Page 9: ...be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth Voltage dips short interruptions and voltage variations on power supply input li...

Page 10: ...device pursuant to part 15 of the FCC rules These limits are designed to provide reasonable protection against harmful interference in a residential installation The product generates uses and can rad...

Page 11: ...r replacements Do not let the sticky side of the pads touch anything including greasy fingertips Cleaning and Maintenance PROGRAM NAME Mode 1 Massage 62 5 100 240 1 500 62 5 100 240 Mode 2 Beat Mode 3...

Page 12: ...ry the pad Is the adhesive surface of the pad damaged Replace the pad It is difficult to attach the pad to the skin Adhesive surface of pads is not sticky Pads get deteriorative Contact the vendor for...

Reviews: