2
data display.
Pressure overload: When the pressure measurement value
exceeds 110%FS, overpressure is displayed and an alarm is
given.
Temperature measurement: (0~50)
℃
; resolution 0.1
℃
;
accuracy: ±0.5
℃
.
Operating environment:
a.
Ambient temperature :(-5~50)
℃
;
b.
Relative humidity:
<
95% (no condensation);
c.
Atmospheric pressure :(86 ~ 106) kPa.
Storage temperature :(-30 ~ 80)
℃
.
Display: 2.8-inch color screen, 5-digit display, Chinese and
English can be switched.
Power supply: built-in 3.7v lithium battery power supply, with
5V power adapter.
Auto power-off function: turn off the auto power-off function
and set the auto power-off time in the system information.
Communication serial port configuration: baud rate: 9600;
check bit: none; data bit: 8; stop bit: 1;
Dimension: header Φ 115 mm x 45 mm; total length: 185 mm.
Weight: about 0.5kg.
Pressure connection: M20×1.5 (can be customized according to
user needs).