background image

 

The material contained in this document is the property of Electronics & Innovation Ltd., it is 
subject to change without notice.  
 
September 2010 

12 

Revision A 

 

d) 

Adjust  the  output  level  of  the  sweep/generator  for  full  vertical 
deflection on the oscilloscope face. 

e) 

Calibrate  the  scope  face  to  show  3  dB  in  1  dB  steps  by 
attenuating the sweep/generator in 1 dB. 

f) 

Return  sweep/generator  output  level  to  full  deflection.  Rotate  the 
step attenuator (CCW) so that the output is reduced by 60 dB. 

g) 

Reconnect the A1000 into the test set-up of Figure  4-1.  

h) 

Place the A1000 power switch to the "ON" position. 

i) 

Observe the gain versus frequency sweep on the oscilloscope. 

 

The average gain  should be greater 60 dB  

 

The gain variation should be within the 3 dB as shown on the 
oscilloscope. 

 
 
 
4.4.2     Measurement of Harmonics  
 

 

 
Figure 4-2. 
 
 
1.)   Equipment Required: 

a) 

Sweep/Signal Generator  

b) 

Calorimetric Power Meter HP435B 

50 dB Attenuator 

Sweep Generator 

RF Output 

300 KHz 

– 35 

MHz 

Spectrum 
Analyzer 

Power Meter 

30 dB 
coupler 

Summary of Contents for A 1000

Page 1: ...ber 2010 Revision A 1 A 1000 Broadband Power Amplifier HIGH RF VOLTAGES MAY BE PRESENT AT THE OUTPUT OF THIS UNIT All operating personnel should use extreme caution in handling these voltages and be t...

Page 2: ...lly operational such that it performs according to the original specifications free of charge to the original purchaser Should warranty service be required the unit must be returned to E I freight cos...

Page 3: ...SSB pulse and any complex modulation signals The amplifier has 60 dB gain it is unconditionally stable and will not oscillate even with combinations of mismatched source and load impedance It is prot...

Page 4: ...conditions without damage POWER OUTPUT METER True RMS power detection 5 of full scale accuracy POWER REQUIREMENTS 3 208 VAC 47 63 Hz 30 Amps SIZE 30 X 24 X 62 inches L x W X H 762 X 210 x 1575 mm L x...

Page 5: ...hieve maximum output Input voltages in excess of 2 volts peak may permanently damage the instrument b The unit should not be driven at an input of over 1 dBm for a period of greater than one minute ii...

Page 6: ...ricted If the fault persists please contact Field Service 2 3 3 RS 232 Interface The A500 features a standard RS 232 serial interface suitable for connection to a PC or host system The communication p...

Page 7: ...BNC port The driver output signal is fed through a length of coaxial cable to the input of the IPA the output is fed to the power amplifier splitter This splitter has four outputs Each output is then...

Page 8: ...t is subject to change without notice September 2010 8 Revision A There are three switch mode power supply modules that provide a 42 VDC 300 ampere source Another provides 24 VDC outputs and provides...

Page 9: ...riodic maintenance The instrument is unconditionally stable and is fail safe under all load conditions Damage can only be externally caused by the incorrect selection of the AC supply voltage or by an...

Page 10: ...nect a sweep generator capable of sweeping the frequency range 300 KHz to 35 MHz to the input connector c Adjust the output level of the sweep generator so that a 50 ohm video detector connected at th...

Page 11: ...D151 e Attenuator 60 dB 1000 Watts Bird Figure 4 1 Gain Measurement 2 Connect the equipment as shown in Figure 4 1 then proceed as follows a Set the oscilloscope to DC Time cm to Ext X and gain to l0m...

Page 12: ...full deflection Rotate the step attenuator CCW so that the output is reduced by 60 dB g Reconnect the A1000 into the test set up of Figure 4 1 h Place the A1000 power switch to the ON position i Obse...

Page 13: ...place in a rigid outer box of wood fiberboard or very strong corrugated cardboard Use ample soft packing to prevent movement Provide additional support for projecting parts to relieve these of unneces...

Reviews: