background image

14 

K8000 000 00EJ 02 065 0A0 17 051 2  

4.1  Working principle 

The gas sample is taken in through the sample probe, by a diaphragm suction pump inside the instrument. 
The measuring probe has a sliding cone that allows the probe to be inserted in holes with a diameter of 11 mm to 
16 mm and to adjust the immersion depth: the sampling point must be roughly in the center of the flue pipe/
stack.
 
The gas  sample is cleaned of humidity and impurities by a condensate trap and filter positioned along the rubber 
hose that connects the probe to the analyzer. 
The gas components are then analyzed by the electrochemical sensors.  
Oxygen  (%O

2

)  is  measured  with  an  electrochemical  cell  that acts  like  a battery  which,  over  time, is  apt  to  lose 

sensitivity. 
The toxic gases (CO, NO) are measured with electrochemical sensors that are not subject to natural deterioration 
being intrinsically lacking of oxidation processes.  
The  electrochemical  cell  guarantees  high  precision  results  in  a  time interval  of  up  to  about  60 minutes  during 
which the instrument can be considered very stable. When measurement is going to take a long time, we suggest 
auto-zeroing the instrument again and flushing the inside of the pneumatic circuit for three minutes with clean air. 
During the zero calibrating phase, the instrument aspirates clean air from the environment and detects the cells’ 
drifts  from  zero  (20.95%  for  the  O2  cell),  then  compares  them  with  the programmed  values  and  compensates 
them. The pressure sen sor autozero must, in all cases, be done manually prior to measuring pressure. 
The  values  measured  and  calculated by  the microprocessor  are  viewed  on  the LCD  display  which is  backlit  to 
ensure easy reading even when lighting is poor. 
 

4.2  Measurement cells 

The measurement  cells  are  electrochemical  cells  made  up  of an anode, a  cathode, and  an electrolytic  solution, 
which  depends  on  the  type  of  gas  to  be  analyzed.  The  gas  penetrates  the  cell  through  a  selective  diffusion 
membrane  and  generates  an  electric  current  proportional  to  the  absorbed  gas.  Such  current  is  measured, 
digitalized, temperature-compensated, processed by the microprocessor, and displayed.  
The  gas  shall  not  be  at  a  pressure  such  to  damage  or  destroy  sensors.  The  maximum  estimated  allowed 
pressure is  ±100hPa gage. 
The response times of the measurement cells used in the analyzer are: 
O

2

  

50 sec. at 90% of the measured value  

CO 

60 sec. at 90% of the measured value 

NO 

=   40 sec. at 90% of the measured value 

It is therefore suggested to wait 5 minutes (anyway not less than 3 minutes) in order to get reliable analysis data.  
If sensors of poison gases are submitted to concentrations higher than 50% of their measurement range for more 
than  10  minutes  continuously,  they  can  show  up  to  ±2% drift  as  well as  a  longer  time  to  return  to  zero.  In  this 
case,  before  turning  off  the  analyzer,  it is  advisable  to  wait  for  the  measured  value  be lower  than  20ppm  by 
disconnecting the probe and letting the pump  run in clean air. 
 

4.3  Connecting the sample probe  

The  sampling  probe  is  made  up  of  an  INOX  steel  tube  with  a  plastic  hand  grip  and  an  internal  K-type 
thermocouple  (Ni-NiCr)  for  measuring  the  fumes  temperature  up  to  800°C.  The  probe  is  connected  to  the 
analyzer  through  a  double  flexible  hose,  a  filter  group  and  a  compensated  cable  for  the  thermocouple.  The 
polarized  connector  of  the  thermocouple  is  to  be  connected  to  the  special  outlet  on  the  lower  side  of  the 
instrument. It is not possible to perform a wrong connection thanks to the different width of contacts. Connect the 
shorter  tube of the probe  to  the  filter group  (fine dust/condensate  trap)  which, in  turn,  shall be  connected  to  the 
central  connector  of  the  instrument  marked  with  letter  “A”.  Connect  the  longer  tube,  ending  with  a  male 
connector,  to  the  negative  pressure  input  of  the  instrument  marked  with  letter  “P-”.  The  different  diameter  of 
connectors does not permit any wrong connections; this permits to avoid any damages to the instrument.  
 

4.4  Condensate trap and fine dust filter 

The sample gas to be analyzed shall reach the measurement cells after being properly dehumidified and purified 
from  the  residual  combustion  products.  To  this  purpose,  a  condensate  trap  is  used,  which  consists  of  a 
transparent  polycarbonate  cylinder  placed  along  the  rubber  hose  of  the  sampling  probe.  Its  purpose  is  to 
decrease  the  air  speed  so  that  the  heavier  fine  dust  particles  can  precipitate  and  the  vapor in  the  combustion 
gases can condensate.  
The  condensate  trap  must  be always  kept  in  the  vertical position in order to prevent  condensate  from  touching 
the measurement  cells. This  is  also  the  reason  why it  is  important  to  periodically drain  the trap,  anyhow at  the 
end of each test (see chapter 'MAINTENANCE'). 
A  replaceable  low-porosity  line  filter  is  placed  after  the  condensate  trap  aimed  at  keeping  the  solid  particles 
su spended  in  the  gases.  It  is  recommended  to  replace  the  filter  whenever  visibly  dirty  (see  chapter 
'MAINTENANCE'). 

4.0  OPERATION 

Summary of Contents for BTU900

Page 1: ...BTU900 BTU900 Portable Combustion Gas Analyzers...

Page 2: ......

Page 3: ...6 Connecting the TcK probe 15 4 7 Keypad overview 16 4 8 Info Menu 17 4 8 1 Flow Chart Info Menu 18 4 9 Analysis configuration menu 20 4 9 1 Flow Chart Analysis configuration menu 22 4 10 Instrument...

Page 4: ...Cleaning the sample probe 52 6 4 Maintaining the water trap filter unit 52 6 5 Replacing the particulate filter 53 6 6 Replacing the gas sensors 53 6 7 On site recalibration 57 6 7 1 Flow Chart On si...

Page 5: ...pre calibrated and BTU900 does not need recalibrating Moreover Operator interface user friendly so much so that it can be used without the instruction manual Luminous and large LCD display easy readab...

Page 6: ...ANUAL FLUE SAMPLING PROBE 300mm 12 10 Dual Hose SMART IINCOMING COMBUSTION AIR TEMPERATURE PROBE Optional WATER TRAP PRESSURE MANOMETER BATTERY CHARGER BATTERY CHARGER CABLE US PLUG MEMORY PC SOFTWARE...

Page 7: ...with 20um particles Suction pump 1 2 l min heads at the flue up to 135hPa Condensate trap Outside the instrument Carbon black Using an optional external Smoke hand pump it is possible to simply enter...

Page 8: ...al pressure H H H H A connector sample probe input by means of the water trap I I I I P connector positive input for measuring draft M M M M Battery charger socket N N N N Serial cable socket for conn...

Page 9: ...of Fig 2 2 Measurement Cells The instrument uses pre calibrated gas sensors of the long lasting FLEX Sensor series for measuring oxygen O2 carbon monoxide CO and nitrogen oxide NO Up to 4 alarms can...

Page 10: ...uels used for calculating losses and efficiencies Smoke measurements It is possible to enter the smoke values measured according to the standard Smoke index scale The instrument will calculate the ave...

Page 11: ...on the type of fuel 2 Stated precision includes error of the external sensor RTD Pt100 class A DIN 43760 1980 3 Stated precision includes error of the external sensor type K thermocouple class 1 IEC58...

Page 12: ...ery will be recharged whilst the instrument is being used The battery charging cycle takes up to 3 hours for a complete charge and finishes automatically ATTENTION If the instrument is not going to be...

Page 13: ...WEEN 90Vac AND 264Vac INPUT FREQUENCY 50 60Hz THE LOW VOLTAGE OUTPUT IS 12 VOLT WITH AN OUTPUT CURRENT GREATER THAN 1 5A LOW VOLTAGE SUPPLY CONNECTOR DC PLUG 2 1x5 5x9 mm WITH CENTRAL POSITIVE AND OUT...

Page 14: ...at 90 of the measured value NO 40 sec at 90 of the measured value It is therefore suggested to wait 5 minutes anyway not less than 3 minutes in order to get reliable analysis data If sensors of poiso...

Page 15: ...water delivery and return temperature by connecting some special probes If temperature is taken on the pipe it is suggested to use arc probes with a suitable diameter ATTENTION KEEP THE CONDENSATE TRA...

Page 16: ...turn on off the instrument it is necessary to press and hold the On Off button for at least 2 seconds Configure Info Menu On Off Measurements Menu Analysis Menu Print Menu Zoom IN Select Modify Zoom...

Page 17: ...sensors thus allowing for a quick troubleshooting in case of issues referable to the sensors Gas path check Tests the tightness of the gas probe pneumatic path Memories diagnostics At instrument turn...

Page 18: ...NSORS TYPE QUIT 1 3 2 O2 NO CO Use arrows to scroll the parameters of each selected cell Here below are the data which can be display ed through the sensors troubleshooting menu Type Ty pe of sensor R...

Page 19: ...ath Check Memories diagnostic INFORMATION Connect to port P Apply cap to probe Press OK to start GAS PATH CHECK Battery status Sensors Config Sensors diagnostic Gas Path Check Memories diagnostic INFO...

Page 20: ...n air temperature as measured from the instrument in case this value is unknow the operator is recommended to enter 50 for this value NOx NO Factor NOx NO all the nitrogen oxides which are present in...

Page 21: ...tor will be printed on the analysis report Report header setup This allows the Company or Owner s name to be entered in four lines with 24 characters each together with other details e g address tel n...

Page 22: ...4 Oil Diesel Wood Pellets 8 FUEL 2 Oil 4 Oil Diesel Wood Pellets 8 Coal FUEL 4 Oil Diesel Wood Pellets 8 Coal Bio Fuel 5 FUEL A1 0 0280 B 0 0090 CO2t 11 70 LHV 51000 kJ kg HHV 56500 kJ kg FUEL PARAME...

Page 23: ...UNITS ppm hPa C ppm CO NO NOx Pressure Temperature MEASUREMENT UNITS ppm hPa C ppm CO NO NOx Pressure Temperature MEASUREMENT UNITS ppm hPa C ppm CO NO NOx Pressure Temperature MEASUREMENT UNITS ppm h...

Page 24: ...e Duration Print AUTOMATIC ANALYSIS auto s 120 manual Mode Duration Print AUTOMATIC ANALYSIS auto s 120 manual Mode Duration Print AUTOMATIC ANALYSIS auto s 120 manual Mode Duration Print AUTOMATIC AN...

Page 25: ...t Unit ALARMS maximum CO 1 1500 ppm Number Measure Active Limit Unit ALARMS minimum CO 1 1500 ppm Modify the v alue of the highlight ed f igure with keys Select alarm number 1 5 with the key s The v a...

Page 26: ...task is done If you wish to modify a letter or a whole line all you need to do is position the cursor in front of the letter to be cancelled by means of the cursor keys in the f irst row of controls...

Page 27: ...t The display contrast may be increased or decreased by acting on cursor keys This operation may be performed even when the introductory screen is active Time Date This allows the current time and dat...

Page 28: ...DATE EU 12 12 12 12 12 Time Date Mode TIME DATE EU 12 12 12 13 12 Bluetooth Calibration Dispaly contrast Time Date Microman meter CONFIGURATION 1 2 3 4 5 6 7 8 Activ ates the Configuration Informatio...

Page 29: ...NFIGURATION Inlet MICROMANOMETER P Inlet MICROMANOMETER P Inlet MICROMANOMETER P Bluetooth Calibration Dispaly contrast Time Date Microman meter INFORMATION 1 2 3 4 5 6 7 8 Dispaly contrast Time Date...

Page 30: ...memory within which to record any effected analyses or other data such as draft smoke and ambient CO NO values When the menu is accessed a preview of all saved data will appear Recall memory This menu...

Page 31: ...no Smoke no CO NO amb no SAVE ANALYSIS SAVE QUIT Save analysis Display average Select memory Recall memory Delete single MEMORY Save analysis Display average Select memory Recall memory Delete single...

Page 32: ...ect memory Recall memory Delete single Delete all MEMORY WARNING Delete all data in the memory 001 DELETE SINGLE QUIT DELETE WARNING Delete all data in the memory 001 Please wait DELETE SINGLE WARNING...

Page 33: ...s sampled in the combustion analysis and when sampled the draft smoke and CO NO ambient gas values Partial only reports the combustion analysis measurement values and information without any header co...

Page 34: ...ull 1 4 12 1 Flow Chart Print Menu Print report Print setup Print test Printer type PRINT Print report Print setup Print test Printer type PRINT Analysis Model PRINT REPORT PRINT QUIT full running Ana...

Page 35: ...code of the printer that must be connected 0000_ TUVXYWZ 012345678 PRINTER PIN Enter the PIN code of the selected printer to complete the pairing procedure TU12_ TUVXYWZ 012345678 PRINTER PIN Use EDIT...

Page 36: ...ons CO2 Carbon dioxide percentage in the Flue gas T Difference between flue gases temperature and combustion supply air temperature NOx Nitrogen oxides concentration in flue gases Loss The sensible ef...

Page 37: ...1p RefO2 Auto 001 Auto 001 O2 4 2 CO2 9 3 Xair 1 25 Tf 190 1C Ta 15 4C T 74 7C Eff 83 4 Loss16 6 CO 146p NO 40p NOX 51p RefO RefO RefO RefO2 2 2 2 Eff 83 4 Loss 16 6 CO 146p NO 40p NOX 51p RefO2 Auto...

Page 38: ...e fuel of the plant being tested Select the memory wherein to store the acquired data By pressing the Analysis key once more and starting from any of the above screens the user may proceed as follows...

Page 39: ...low Chart Draft Menu Bef ore starting the pressure zeroing sequence pay attention to remove the gas probef rom the stack Connect the probe pressure input hose to the instrument P input WARNING Use P p...

Page 40: ...Omax 25 ppm Recommended exposure limit REL stipulated by the National Institute for Occupational Safety and Health NIOSH equivalent to 30 mg m3 and calculated as an 8 hour Time Weighted Average TWA AT...

Page 41: ...T PRINT COmax NOmax AMBIENT CO NO START START START START KEEP PPM PPM 0 0 0 01hPa PRESSURE ZERO ZERO ZERO ZERO HOLD HOLD HOLD HOLD 0 00hPa PRESSURE ZERO ZERO ZERO ZERO HOLD HOLD HOLD HOLD 0 00hPa PRE...

Page 42: ...end Adjusts the display contrast Or automatically after 10 seconds Natural gas L P G Diesel oil 2 Fuel oil Methane FUEL 1 2 3 OPERATOR ID Copies Model PRINT SETUP full 1 Mode AUTOMATIC ANALYSIS manua...

Page 43: ...ent and screwing in the positioning cone provided with the probe in this way no air is drawn from the outside during sampling The screw on the cone allows the probe to be stopped at the right measurin...

Page 44: ...is If on the other hand manual analysis mode is chosen flue gas analysis will proceed manually please see relative Flow Chart In this case the print settings and automatic test duration will not be co...

Page 45: ...1 Mode AUTOMATIC ANALYSIS manual RefO2 Auto 001 O2 4 2 CO2 9 3 Xair 1 25 Tf 190 1C Ta 15 4C T 74 7C Eff 83 4 Loss16 6 CO 146p NO 40p NOX 51p WARNING Insert the gas probe in the chimney AUTOZERO STARTE...

Page 46: ...Printer type PRINT Copies Model PRINT SETUP full 1 Print report Print setup Print test Printer type PRINT RefO2 O2 4 2 CO2 9 3 Xair 1 25 Tf 190 1C Ta 15 4C T 74 7C Eff 83 4 Loss16 6 CO 146p NO 40p NOX...

Page 47: ...2 CO2 9 3 Xair 1 25 Tf 190 1C Ta 15 4C T 74 7C Eff 83 4 Loss16 6 CO 146p NO 40p NOX 51p RefO2 O2 4 2 CO2 9 3 Xair 1 25 Tf 190 1C Ta 15 4C T 74 7C Eff 83 4 Loss16 6 CO 146p NO 40p NOX 51p RefO2 O2 4 2...

Page 48: ...15 4C T 74 7C Eff 83 4 Loss16 6 CO 146p NO 40p NOX 51p RefO2 001 1 120 O2 4 2 CO2 9 3 Xair 1 25 Tf 190 1C Ta 15 4C T 74 7C Eff 83 4 Loss16 6 CO 146p NO 40p NOX 51p RefO2 001 2 120 O2 4 2 CO2 9 3 Xair...

Page 49: ...iezoresistive transducer to measure positive and negative pressures This sensor which is mounted on the instrument is of the differential type If the special KIT is purchased the sensor can be used to...

Page 50: ...re measuring accuracy recalibration can only be performed by a qualified E Instruments Service Center Chart 5 4 illustrates the characteristics inherent to each sensor 5 4 Gas sensors life table Notes...

Page 51: ...expandable instrument is arranged in a way to accept one sensor in positions S3 Identify with the help of paragraph 5 2 Sensor types and relevant positioning the sensor which must be added to the exis...

Page 52: ...ete overhaul and thorough internal cleaning E Instruments highly qualified staff is always at your disposal and will provide you with all the sales technical application and maintenance details requir...

Page 53: ...he gas sensors The gas sensors of the instrument shall be periodically replaced see the following table with new or recalibrated sensors The user can easily perform this replacement operation accordin...

Page 54: ...12 Electrical connection 3 Locate the sensor to be replaced here is an example of a connected sensor to be replaced 4 Disconnect the sensor to be replaced here is an example of a disconnected sensor t...

Page 55: ...nnected to its socket rotate it counter clockwise to remove it Here is an example of a rotated sensor ATTENTION While rotating the sensor take care not to exert any pressure onto the printed circuit a...

Page 56: ...an external polarisation battery the settling time is reduced down to 2 hours CODE MEASURED GAS POSITION SETTLING TIME FLEX Sensor O2 Cod AAC SE15 O2 Oxygen S1 2 hours 1 FLEX Sensor CO Low H2 Cod AAC...

Page 57: ...on When depressed for at least 2 seconds turns the instrument on RefO2 Auto 001 O2 4 2 CO2 9 3 Xair 1 25 Tf 190 1C Ta 15 4C T 74 7C Eff 83 4 Loss16 6 CO 146p NO 40p NOX 51p WARNING Insert the gas prob...

Page 58: ...n screenshot information about the calibration in use and sensor output are displayed Action selection of action to make calibrate save new calibration set original bring back factory original calibra...

Page 59: ...ied 1022 0 Applied 1022 0 Applied 1022 0 P P P P Measured 990 5 P Is 82 22 uA Ia 10 17 uA Status original CO CALIBRATION With keys set the cylinder concentration in the Applied row Action calibrate Ac...

Page 60: ...tions to replace the battery pack 1 Undo the screw on the battery compartment cover and remove it 2 Remove the battery pack Battery pack connector 3 Remove the battery pack connector and replace the p...

Page 61: ...m Send the instrument to a service center The analysis screen gives a flue gas temperature Tf error a The thermocouple is not connected connect the thermocouple to the analyzer b The sensor has been e...

Page 62: ...to store the instrument at higher temperatures b The battery pack is old Battery capacity tends to diminish with age If battery life has become unacceptable replace the battery pack The values shown...

Page 63: ...extended tempe rature range with 3 mt cable AAC SF22A 300 mm gas probe 1100 C extended tempe rature range with 3 mt cable AAC SF25A 750 mm gas probe 1100 C extended tempe rature range with 3 mt cable...

Page 64: ......

Reviews: