11
K8000 000 00EJ 02 065 0A0 17 051 2
Probe for measuring the ionization current
With this special probe it is possible to measure the ionization current of a boiler and check its value depending
on the boiler’s technical features.
Calibration certificate
The instrument is calibrated by comparing to National Calibration Standards provided by a Metrology Lab.,
certified periodically by internationally recognized laboratories.
A calibration certificate is provided with each and every instrument where every parameter is accompanied by the
relevant nominal value, measured value, permissible error tolerances and measured error.
Electromagnetic compatibility
The instrument was designed to comply with Council Directive 2004/108/EC governing electromagnetic
compatibility.
2.3 Measurement Ranges and Accuracies
All data relative to concentration accuracies are referred to an instrument operating at a constant temperature
within the correct operating range (-5°C .. +45°C), being in operation for at least 15 minutes, powered by its
internal battery and after completion of auto-zero procedure.
Notes: (1) The maximum CO
2
value displayed depends on the type of fuel.
(2) Stated precision includes error of the external sensor RTD Pt100 class A DIN 43760 (1980).
(3) Stated precision includes error of the external sensor type K thermocouple class 1 IEC584.
(4) Pressures greater than 300 inH
2
O may permanently damage sensors or impair their characteristics.
1
2
3
4
5
6
7
8
MEASUREMENT
SENSOR
RANGE
RESOLUTION
ACCURACY
O
2
Electrochemic al sensor
0 .. 25. 0% vol
0.1% vol
±0. 2% vol
CO
W ith low sensitivity to
H2
Electrochemic al sensor
0 .. 8000 ppm
1 ppm
±20 ppm 0 .. 400 ppm
±5% measured val ue 201 . . 4000 ppm
±10% meas ured value 4001 . . 8000 ppm
NO
Electrochemic al sensor
0 .. 5000 ppm
1 ppm
±5 ppm 0 .. 100 ppm
±5% measured val ue 101 .. 5000 ppm
NOx
Calculated
CO
2
Calculated
0 .. 99. 9% vol
0.1% vol
Air temperatur e
Pt100 s ens or
-20.0 .. 120.0 °C
0.1 °C
±0. 5 °C
Flue gas temper ature
TcK sens or
-100.0 .. 1250.0 °C
0.1 °C
±0. 5 °C 0 .. 10 0 °C
±0. 5% meas ured value 101 .. 1250 °C
Pressure
(draft & differential)
Piezoel ectric sens or
-4.014 .. 80.292 inH
2
O
0.004 inH
2
O
±1% measured val ue -4. 014 .. 0.806 inH
2
O
±0. 008 inH
2
O -0.802 .. 0.802 i nH
2
O
±1% measured val ue 0.806 . . 0.292 inH
2
O
Differential temp erature
Calculated
0 .. 1250. 0 °C
0.1 °C
Air index
Calculated
0.00 .. 9. 50
0.01
Excess ai r
Calculated
0 .. 850 %
1 %
Stack loss
Calculated
0.0 .. 100.0 %
0.1 %
Efficiency
Calculated
0.0 .. 100.0 %
0.1 %
Efficiency
(condensing)
Calculated
0.0 .. 120.0 %
0.1 %
Smoke index
External instrument
0 .. 9
(1)
(2)
(4)
(3)