Measurements 
Ambient CO 

CO 

ppm 

CO max 

ppm 

It  is  mandatory  to  perform  the 
instrument  autozero  in  fresh,  clean 
outdoor air.

 

7/08/14 

10:00 

Measurements 
Ambient CO 

CO 

ppm 

CO max 

ppm 

Connect the gas probe to the instrument and perform 

the measurement. 

Wait 5 minutes.

 

7/08/14 

10:00 

Measurements 
Draft 

P diff 

inH2O 

0.000 

 

 

 

7/08/14 

10:00 

Measurements 
Draft 

P diff 

inH2O 

 - 0.020 

 

 

 

Once the 

pressure zeroing 

is completed 

insert the probe in 

the chimney and 

measure the draft. 

AA SF--A 

AAC TA03A 

AAC KP01 

AA SA08 

7/08/14 

10:00 

Measurements 
Draft 

P diff 

inH2O 

 0.001 

 

 

 

Before zeroing the 

pressure it is 
mandatory to 
REMOVE the 

probe from the 

chimney. 

 

View & Save the 

DRAFT 

measurement. 

ZERO Draft 

Sensor 

 

Save the 

measurement 

 

Draft 

Ambient CO 

7/08/14 

10:00 

Eff. tot 

91.4 

Loss tot 

8.6 

ΔT 

°F 

74.7 

T air 

°F 

15.4 

T flue 

°F 

190.1 

Ex. Air 

 

1.25 

CO

2

 

9.3 

O

2

 

4.2 

Combustion analysis 
UNI 10389 

KEEP 

PRESSED 

FOR 2 

SECONDS 

UNTIL IT 

BEEPS 

Insert  the  gas  probe  in  the 
chimney.  The  gas  autozero 
w i l l  

b e  

a u t o m a t i c a l l y 

performed  thanks  to  the 
internal diverting electrovalve. 

Memory 
Select 

7/08/14 

10:00 

Customer

 

E Instruments 

Memory 

Address

 

402 Middletown  

Ste 216 

Phone

 

215-750-1212 

Langhorne, PA 19047 

Boiler

 

xxxx 

Date 

7/08/14 

In  "Select"  can  be 
inserted 

all 

the 

customer data. 

Memory 
Data logger 

7/08/14 

10:00 

Samples 

10 

Mode 

Data logger 

Interval 

ppm 

60 

Print 

auto 

Memory 

auto 

In  "Data  logger",  the 
user  can  define  the 
anal ysis, 

memory 

selection  and  printing 
modes. 

7/08/14 

10:00 

Save 

Select 

Delete 

Average 

Data logger 

Usage % 

Memory 

OK 

OK 

Data logger 

Select 

WARNING! 

 

Be sure all connections are tight to assure accurate sampling. 

 

It is important that water trap & filter be installed vertically to 

prevent damage from moisture & particles to sensors. 

 

When conducting measurements,  the water trap/filter 

assembly MUST be in a VERTICAL position. 

 

When  testing  is  completed,  always  drain  the  water 

trap with any condensation (after EVERY test!). 