Display

................................Graphic display 128x128 pixels

Dim. 50 x 54 mm, blue blacklit, Display of 6 

measurements (including 4 simultaneously)

Housing

..............................IP54, ABS shock-proof

Keypad

................................Metal-coated, 5 keys, 1 joystick 

Conformity

.........................Electromagnetical compatibility

(NF EN 61326-1 norm)

Power supply

.....................4 alkaline batteries 1.5V LR6

Operating environment

......Neutral gas

Operating temperature

......from 0 to  50°C

Storage temperature

..........from -20 to +80°C

Auto shut-off

......................adjustable from 0 to 120 min

Weight

.................................380g

Languages

..........................English, French 

4

Technical features

Specifications

I – Technical specifications

Sensing elements

AQ200 connection (see p.6)

Air quality probe
CO

2

 :

 NDIR infrared sensor (Non dispersive- infrared)

CO : 

Electrochemical sensor

Temperature : 

Pt100 class A 

Hygrometry :

 capacitive hygrometry sensor

Climatic conditions module

Hygrometry :

 capacitive hygrometry sensor

Temperature : 

semiconductor temperature sensor

Air pressure :

 capacitive sensor

Thermocouple probes :

 type K, J and T class 1

Pt100 Smart-plus probes :

 Pt100 class 1/3 Din

Measuring units

Measuring ranges

Accuracy*

Resolutions

V, mA

from 0 to 2.5 V
from 0 to 10 V
from 0 to 4/20 mA 

±2mV

±10mV

±0.01mA

°C, °F

ppm

from -20 to +80°C

±0.4% of reading ±0.3°C

0.1 °C

from 0 to 100 ppm
from 101 to 1,000 ppm

±5 ppm

1 ppm

CURRENT / VOLTAGE

CO / Temperature

from -200 to 1300°C
from -100 to 750°C
from -200 to 400°C

±1,1°C ou ±0.4% Reading value***

±0.8°C ou ±0.4% Reading value***

±0.5°C ou ±0.4% Reading value***

0.1 °C
0.1 °C
0.1 °C

°C, °F

K :

J :

T :

THERMOCOUPLE  (See related datasheet)

CLIMATIC CONDITIONS

Hygro.

Air pressure

%RH

from 5 to 95%RH

Accuracy** (Repeatability, linearity, 
hysteresis) 

:  ±1.8%RH (from 15°C to 25°C)

Factory calibration uncertainty:

 ±0.88 %RH

Temperature dependence : 

±0.04 x (T-20) %RH (if T<15°C or T>25°C)

0.1 %RH

from 800 to 1,100 hPa

±3 hPa

1 hPa

hPa

Temp.

from -20 to +80°C

±0.4% of reading ±0.3°C

0.1 °C

°C, °F

Pt100 Smart-plus or wireless probes (See related datasheet)

from -50 to 250°C
(According to model)

±0.3% of reading ±0.25°C
(According to model)

0.01 °C

°C, °F

°C, °F

ppm

from -20 to +80°C

±0.4% of reading ±0.3°C

0.1 °C

from 0 to 5,000 ppm

± 3% of reading ±50 ppm

1 ppm

CO

2

 / Temperature

°C, °F

ppm

%RH

from -20 to +80°C

±0.4% of reading ±0.3°C

0.1 °C

from 0 to 5,000 ppm
from 5 to 95%RH

±3% of reading  ±50 ppm

1 ppm

CO

2

 / Temperature / Hygrometry

See Climatic conditions module

0.1 %RH

CO

Temp.

CO

2

Temp.

CO

2

Temp.

Hygro.

0.001 V
0.01 V
0.01 mA

+

*All accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurements carried out in the same conditions, or carried out with required compensation.

 

** As per NFX 15-113 and the Charter 2000/2001 HYGROMETERS, GAL (Guaranteed Accuracy Limit) which has been calculated with a coverage factor value of 2 is ±2.88%RH between 18 and 28°C on the measuring range from 5 to 95%RH. Sensor drift is less than 1%RH/year.

***The accuracy is expressed either by a deviation in °C, or  by a percentage of the value concerned. Only the bigger value is considered.

±3% of reading ±5ppm

Summary of Contents for AQ 200

Page 1: ...AQ 200 AQ 200 Air Quality...

Page 2: ...Using wire probes and modules 8 Use wireless probes 8 Functions 8 Hold Min Max 8 Configuration 9 Thermocouple type 9 Display 9 Units 9 Delta T 9 Alarms 9 Recording 9 Parameters 11 Language 11 Date tim...

Page 3: ...0 4 Reading value 0 5 C ou 0 4 Reading value 0 1 C 0 1 C 0 1 C C F K J T THERMOCOUPLE See related datasheet CLIMATIC CONDITIONS Hygro Air pressure RH from 5 to 95 RH Accuracy Repeatability linearity...

Page 4: ...5 Description II Introduction...

Page 5: ...the instrument Mini Din C1 connector Mini Din C2 connector Probes are connected on min DIN connectors C1 and or C2 Wireless probe instrument communication Wireless communication between probe and inst...

Page 6: ...tion with right and left keys Connections can be activated or deactivated with or Infos Measure Params Select a sub menu with access keys or with arrow keys Measurement Communication interrupted Check...

Page 7: ...re Sensor type Serial number Date of last calibration or adjustement Probes Status enabled ou disabled On enabled mode the probe is connected the measurement is carried out and the value is displayed...

Page 8: ...equired with and Confirm with OK Click on Esc to return to previous screen Delta T When two PT100 probes or 2 thermocouple temperature probes are connected AQ200 can calculate Delta temperature value...

Page 9: ...r can enter a theorical value and a tolerance for the parameter to be controlled Planification must be made via the software a Click on OK or to enter into sub function b Select Planned with and Confi...

Page 10: ...ct the delay in minute Click on OK or to enter into the sub function Select with and OFF in order to disable the automatic shut off or enter the delay from 15 to 120 minutes Confirm wih OK Date time B...

Page 11: ...After Sales Service required for appraisal Maintenance Warranty Info menu This menu allows to view the serial number of instrument and firmware version Battery When battery indicator flashes it is re...

Reviews: