Preliminary operation manual EE600 Differential Pressure Sensor
3 Product Description
FOR CONDUIT
INSTALLATION
CABLE GLAND M16x1.5
90
±0.3
mm
3.54
±0.11
“
46mm
1.81“
5mm
0.2“
80.6mm 3.17“
5mm
0.2“
Pressure fittings Ø 5mm
+ : High pressure
- : Low pressure
Fig. 1
Dimensions
1
2
3
5
4
1. Pressure connection nipples
2. DIP switches
3. Spring terminals
4. Push buttons for zero point and
span point adjustment
5. Status LED
Fig. 2
Design and components
15...35 V DC
24 V AC ±20 %
Fig. 3
Connection diagram
4 Initial Setup
Each EE600 leaves the E+E factory with the default setup (all switches on „0“):
• Measurement range:
100% of the full scale (1000 Pa or 10000 Pa)
• Output signals:
0-10V and 4-20mA
• Time constant:
50ms
• Backlight display:
on
EE600 is fully configurable. Before commissioning, set the individual configuration of EE600 according
to the application requirements using the DIP switches S1 to S8 as described below. The function of the
DIP switches is indicated also on the interior of the EE600 front cover.
Fig. 4
DIP switches