23
4.3 Configuration adapter
The digital interface is used only for device configuration and customer adjustment.
The configurator software and the drivers are available for download free of charge from our
website
http://www.epluse.com/en/service-support/download-center/
.
The EE300Ex in combination with the configurator software is only permitted for use outside
the hazard area.
It may only be used the cable with protective circuit (aluminum housing on the
8 pin connector).
Connection from CH1 is NOT required for communication.
RS232
(the external power supply required is included as standard with the configuration adapter)
:
USB:
If the upper housing section of the EE300Ex with the electronics and measurement probe is
removed, the lower housing section with the blanking plate (HA011401) should be sealed to
protect against dirt and electrostatic charge.
4.4 Calibration of the current loop
Only appropriately approved multimeters are permitted for calibration in hazardous areas.
During measurement with these multimeters, ensure that the requirements of the system
description (proof of intrinsic safety) remain in force.
PC
PC
PC
PC
Connection cable - HA011061
Configuration adapter -
EE-PCA
4.2 Terminal assignment EE300Ex
Erde
CH1 +
CH1 -
CH2 +
CH2 -