User Manual EE260 Heated Humidity and Temperature Probe for Meteorological Applications

The steps recommended by the manufacturer for installation, inspections and maintenance work must 

be observed and carried out for the safety of the user and for the functionality of the equipment.

Unauthorized product modification leads to loss of all warranty claims. This may be accomplished only 

with an explicit permission of E+E Elektronik

®

!

1.2.3  Mounting, Start-up and Operation

The EE260 humidity and temperature probe for meteorological applications has been produced under 

state of the art manufacturing conditions, has been thoroughly tested and has left the factory after 

fulfilling all safety criteria. The manufacturer has taken all precautions to ensure safe operation of the 

device. The user must ensure that the device is set up and installed in a manner that does not have a 

negative effect on its safe use. The user is responsible for observing all applicable safety guidelines, 

local and international, with respect to safe installation and operation on the device. This user manual 

contains information and warnings that must be observed by the user in order to ensure safe operation.

 

ƒ

Mounting, start-up, operation and maintenance of the device may be performed by qualified staff 

only. Such staff must be authorized by the operator of the facility to carry out the mentioned activities.

 

ƒ

The qualified staff must have read and understood this user manual and must follow the instructions 

contained within.

 

ƒ

All process and electrical connections shall be thoroughly checked by authorized staff before putting 

the device into operation.

 

ƒ

Do not install or start-up a device supposed to be faulty. Make sure that such devices are not 

accidentally used by marking them clearly as faulty.

 

ƒ

A faulty device may only be investigated and possibly repaired by qualified, trained and authorized 

staff. If the fault cannot be fixed, the device shall be removed from the process.

 

ƒ

Service operations other than described in this user manual may only be performed by the 

manufacturer. 

1.3  Environmental Aspects

Products from E+E Elektronik® are developed and manufactured in compliance with all relevant  

environmental protection requirements. Please observe local regulations for the device disposal.

For disposal, the individual components of the device must be separated according to local recycling 

regulations. The electronics shall be disposed of correctly as electronics waste.

2  Scope of Supply

 

ƒ

EE260 – Heated Humidity and Temperature Probe for Meteorological Applications

 

ƒ

Inspection certificate according to DIN EN 10204-3.1

 

ƒ

Quick guide

3  Product Description

3.1  General

The EE260 probe is optimized for accurate and reliable relative humidity (RH) and temperature (T) 

measurement in meteorology and demanding outdoor applications.

Mounted in its upright position, the air first passes over EE260‘s Pt100 temperature sensing element, 

then over the heated probe shaft and over the heated humidity sensing element. This dual heating 

system prevents condensation or icing of the RH sensing element, on the probe head and on the filter 

cap. This leads to short response times and fast recovery as well as to precise RH measurement even 

under continuously high humidity and condensing conditions.

Summary of Contents for EE260

Page 1: ...BA_EE260 v1 1 Modification rights reserved User Manual EE260 Heated Humidity and Temperature Probe for Meteorological Applications...

Page 2: ...equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmf...

Page 3: ...and Adjustment 9 5 1 EE PCS Product Configuration Software 9 5 2 Digital Interface RS485 with Modbus RTU Protocol 9 5 3 Modbus Register Map 11 5 4 Modbus Message Example 12 6 Maintenance and Service...

Page 4: ...Safety Instructions 1 2 1 General Safety Instructions The device and mainly the filter cap shall not be exposed to unnecessary mechanical stress When replacing the filter cap make sure not to touch t...

Page 5: ...connections shall be thoroughly checked by authorized staff before putting the device into operation Do not install or start up a device supposed to be faulty Make sure that such devices are not acci...

Page 6: ...umidity related parameters Dew point temperature Td Frost point temperature Tf Wet bulb temperature Tw Water vapour partial pressure e Mixing ratio r Absolute humidity dv Specific enthalpy h The RH an...

Page 7: ...cho Puls 25 18 25 Fig 2 Dimensions in mm of optional radiation shield ordering code HA010511 3 3 Electrical Connection Pin number Function Wire colors for accessories Connection cable HA010322 23 24 2...

Page 8: ...he HA010511 data sheet for more information Fig 3 EE260 mounted within the optional radiation shield HA010511 4 2 Wall Mount Pole Mount The probe may be installed on a wall or a pole with the help of...

Page 9: ...nd install it on the PC 2 Connect the EE260 to the PC using the Modbus configuration adapter and the EE260 configuration cable 3 Start the EE PCS software 4 Follow the instructions on the EE PCS openi...

Page 10: ...ction code 0x06 Slave ID Modbus address 1 0x00 Modbus protocol settings3 2 0x01 1 Register number starts from 1 2 Protocol address starts from 0 3 For Modbus protocol settings see Application Note Mod...

Page 11: ...d C 1105 0x450 F 1107 0x452 K 1147 0x47A Wet bulb temperature Tw C 1109 0x454 F 1111 0x456 K 1145 0x478 Absolute humidity dv g m 1113 0x458 gr ft 1115 0x45A Mixing ratio r g kg 1121 0x460 gr lb 1123 0...

Page 12: ...sure above water ew mbar 100 4111 0x100E psi 100 4112 0x100F Saturation vapor pressure above ice ei mbar 100 4113 0x1010 psi 100 4114 0x1011 Ice bulb temperature Ti C 100 4119 0x1016 F 100 4120 0x1017...

Page 13: ...the user see the E E cleaning instructions 6 2 Filter Replacement In a dusty or polluted environment it might be necessary to replace the filter cap or the filter membrane respectively once in a whil...

Page 14: ...emperature Please contact your sales representative for assistance Please note When the calibration adjustment procedure with the help of EE PCS starts both heating modes of the EE260 are disabled aut...

Page 15: ...ure with the help of EE PCS starts both heating modes of the EE260 are disabled automatically if they have on status At the end of the procedure the heating comes back to on automatically The temperat...

Page 16: ...EE260 configuration cable1 HA011020 E E Product Configuration Software Download www epluse com configurator EE PCS M12x1 cable connector 8 pole socket HA010704 Connection cable 8 pole M12x1 socket fre...

Page 17: ...ID 235 General Supply voltage 7 30 V DC Power consumption typ 300 mW 25 mA 12 V DC heating included Electrical connection M12x1 8 poles stainless steel 1 4404 Filter PTFE membrane stainless steel body...

Page 18: ...lin 69130 Ecully Tel 33 4 74 72 35 82 E mail info epluse fr E E Elektronik Germany Sch ne Aussicht 8 C 61348 Bad Homburg Tel 49 6172 13881 0 E mail info epluse de E E Elektronik Italy Via Alghero 17 1...