34
After the one’s digit position has also been changed to the desired value, the next pressing of the
Enter key returns the transmitter to the alternating display of Change Zero. Since changing of the
zero has just been completed, press the Next key to proceed to the next menu selection, Change
Full Scale.
5.6
C
HANGE
F
ULL
S
CALE
(U
PPER
R
ANGE
V
ALUE
)
The code
9300
corresponds to selection
CHANGE FULL SCALE
, or Upper Range Value (URV). To
change the full scale value press
ENTER
. The procedure for selecting Plus or Minus is identical to
that described for changing the zero, with the code 9301 corresponding to a plus (+) number and
the code 9302 corresponding to a minus (-) number. The procedure for changing each of the digit
positions is identical to that described for changing the zero. Once the steps of changing the FULL
SCALE have been completed and the
ENTER
key is pressed at the end of the procedure, the display
returns to CHANGE FULL SCALE. Press
NEXT
for the next function
SELECT SENSOR FAIL
SAFE DETECTION
.
5.7
S
ELECT
S
ENSOR
F
AIL
S
AFE
D
ETECTION
The code
9400
corresponds to selecting the Sensor Fail Safe detection. If one desires to change the
Sensor Fail Safe detection then press
ENTER
. The present status of the Sensor Fail Safe is displayed.
It is recommended that one turns off the Sensor Fail Safe when using the TX180 with an input
simulator. It should then be turned on when reconnecting the transmitter to the actual sensor.
The code
9401
indicates that the Sensor Fail Safe detection is turned on.
The code
9402
indicates that the Sensor Fail Safe detection is turned off. When the desired Fail Safe
condition is displayed, pressing the
ENTER
key will change to the new setting and the screen
returns to the
SELECT SENSOR FAIL SAFE
display, code 9400. Pressing the
NEXT
key will then
00
40
004
0
9300
9400
9401
9402
Summary of Contents for TX180
Page 2: ......
Page 26: ...26 Fig 7 Dynisco TX180 ITX190 Configuration Flowchart LI 2 Two Line Display Keypad...
Page 40: ...40 Fig 8 Dynisco TX180 ITX190 Configuration Flowchart LI 2 One Line Display Keypad...
Page 50: ...NOTES...
Page 52: ...DYNISCO INSTRUMENTS 38 FORGE PARKWAY FRANKLIN MA 02038 ATTN MARKETING DEPT Place Stamp Here...