TX180 Temperature Transmitter
21
To change the full scale value press
ENTER
. The procedure for selecting Plus or Minus is identical
to that described for changing the zero. Similarly, the procedure for changing each of the digit
positions is identical to that described for changing the zero. Once the steps of changing the Full
Scale have been completed and the
ENTER
key is pressed at the end of the procedure, the display
returns to
CHANGE FULL SCALE
. Press
NEXT
for the next function
SELECT SENSOR FAIL
SAFE DETECTION
.
4.7
S
ELECT
S
ENSOR
F
AIL
-S
AFE
D
ETECTION
If you want to change the
SENSOR FAIL SAFE
detection press
ENTER
. The present status of the
Sensor Fail Safe is displayed. It is recommended that one turns off the Sensor Fail Safe System when
using the TX180 with an input simulator. It should then be turned on when reconnecting the
transmitter to the actual sensor.
When the desired Fail Safe condition is displayed, pressing the
ENTER
key will change to the new
setting and the screen returns to the
SELECT SENSOR FAIL SAFE
display. Pressing the
NEXT
key will bring up the Fail Safe Reporting selection screen.
4.8
S
ELECT
F
AIL
S
AFE
R
EPORTING
Fail Safe reporting allows the transmitter to change the 4-20mA loop to indicate a failure condition.
This failure may be a sensor failure or a transmitter failure. In any event, the user may select to drive
the loop to 23.0mA, corresponding to the “HI” selection; or 3.6mA, corresponding to the “LO”
selection or to turn the function “OFF”.
SELECT
SENSOR
FAIL
SAFE?
SELECT
FAIL
SAFE
REPORT?
ON
OFF
HIGH?
OFF?
LOW?
Summary of Contents for TX180
Page 2: ......
Page 26: ...26 Fig 7 Dynisco TX180 ITX190 Configuration Flowchart LI 2 Two Line Display Keypad...
Page 40: ...40 Fig 8 Dynisco TX180 ITX190 Configuration Flowchart LI 2 One Line Display Keypad...
Page 50: ...NOTES...
Page 52: ...DYNISCO INSTRUMENTS 38 FORGE PARKWAY FRANKLIN MA 02038 ATTN MARKETING DEPT Place Stamp Here...