17
Get/Set
WarnTripPointHigh
UINT
Valid Range:
(WTPH) If WarnEnable
is set and PV >= WTPH,
(WTPL + WarmHysteresis) <=
TripWarnHigh is set
WTPH <=
(see Attr. 05)
(ATPH - AlarmHysteresis)
TripWarnHigh is cleared
once PV <= WTPH –
WarnHysteresis (see Attr 19).
18
Get/Set
WarnTripPointLow
UINT
Valid Range:
(WTPL)If AlarmEnable is
set and PV <= WTPL,
(ATPL + AlarmHysteresis) <=
TripWarnLow is set
WTPH <= WarnHysteresis
(see Attr. 05)TripWarnLow
is cleared once PV >= WTPL
+ WarnHysteresis (see Attr 19).
19
Get/Set
WarnHysteresis
UINT
Valid Range:
The amount by which
PV must recover to clear
WarnHysteresis <=
either TripWarnHigh
(WTPH - WTPL)
or TripWarnLow.
20
Get/Set
WarnSettlingTime
UINT
Units are in milliseconds
The time that the PV
must exceed the trip
point before TripWarnHigh
or TripWarnLow is set.
21
Get
OverRange
UINT
If PV > OverRange,
then OverRangeFault
is set (see Attr. 03).
22
Get
UnderRange
UINT
If PV < UnderRange,
then UnderRangeFault
is set (see Attr. 03).
25
Set
Execute Rezero
SINT
When set tol, unit
is rezeroed