21
Dymax BlueWave® 75 User Guide
■
The condition of the Lightguide will also affect the intensity. All Lightguides degrade with time, but
intensity will also drop if the Lightguide is bent or kinked. The intensity reading from the Lightguide
should be compared to the intensity reading from a Lightguide Simulator to determine the efficiency.
■
The Bulb is not installed correctly. Bulb orientation is vital during the installation of the Bulb. Refer to the
Bulb installation instructions supplied with all new 75W Bulbs for the proper installation procedure.
Q.) My Footswitch is not operating.
■
Check the connection of the Footswitch into the unit. Check that the Lightguide is fully installed and that
the “Lightguide Seated” indicating light is on.
Q.) What causes my Shutter to hesitate to open or close?
■
Maintaining clean equipment and a clean working environment will help prevent the build up of dust and
other debris from collecting on the mechanical parts of the
BlueWave 75
. Debris that settles in the piston
well (Adjustment Screw) can hinder the movement of the Shutter Solenoid Piston.
■
Shutter alignment to Reflector Mount incorrect.
■
The Shutter is a mechanical part which may wear after extended use.
■
The alignment of the Shutter to the Bulb Mount Assembly is incorrect.
Q.) Why does my
BlueWave 75
seem to run very hot?
■
Replace the Fan Filter Media on the Exhaust Fan and vent located in the back of the equipment. This is
your first line of defense against airborne dust and debris. This Fan Filter Media is supplied with new
Bulbs and should be changed regularly.
■
Ideal operation of this equipment suggests at least 12" of clearance behind the unit for proper ventilation.
Confirm that the Intake Fan is not feeding from the exhaust of other equipment.
■
Confirm that the cooling fan is operating.
■
The unit may already be full of dust and debris and over heating the internal electronics.