Dwyer Instruments RTI2 Series Instructions Manual Download Page 12

Page 11

RTI2 INSTRUCTIONS

Dwyer Instruments, Inc. • 219.879.8868 • www.dwyer-inst.com 

Problem

Probable Causes

Things to try…

Display blank

No power to the unit

Check for minimum 12 Vdc at power terminals

Short in sensor circuit

Disconnect sensor, see if display returns (zero flow 

rate)

Display is in sleep mode

Push any button to reactivate display. (Display goes to 

sleep after about 3 minutes of non-use.)

Display missing pixels

Damaged display module

Contact distributor for return/replacement

Display showing meaningless 

characters

Unit's microcontroller crashed

Disconnect and reconnect power. If problem repeats, 

contact distributor for return/replacement.

Display reads normally but flow 

rate incorrect

Wrong K-factor or time base entered

Enter correct K-factor from meter, fitting, or manual

Display reads normally but 

incorrect pulse output

Wrong pulse output setting

Use "Set P" to correct pulse output setting

Polarity reversed on pulse output 

terminals

Reverse leads

Display reads normally, but no (or 

incorrect) 4-20mA output (RTI2 

only)

Wrong 4mA setting or wrong 20mA 

setting

Use "Set 4" to correct target minimum flow rate. 

Use "Set 20" to correct target top flow rate.

Inadequate loop power supply voltage

Check voltage (For 4-20mA applications, 24 Vdc is 

recommended)

Polarity incorrect in 4-20mA loop 

circuit

Compare to Connections diagram

Display reads zero when there is 

flow

Flow  sensor failed

Consult flow sensor manual for how to test

Break in flow sensor circuit

Check for continuity with multimeter

Flow sensor not battery-compatible

Check flow sensor model number for "micropower 

option"

Display reads flow rate when 

there is none

Long flow sensor wire, running 

parallel to power wires

Reroute wire or change to shielded wire

Flow sensor malfunction

See flow sensor manual to check

Flow "jitter" (oscillating slosh) reads 

as flow

Consult factory for "anti-jitter" setting

Totalizer does not always appear 

to display the total flow

Break in power to meter

The totalizer's memory is only updated every 15 

minutes. If power is lost, the totalizer will retain the 

value last written but will not be updated to reflect any 

flow between the last write and the time the power 

was lost.

TROUBLESHOOTING

Summary of Contents for RTI2 Series

Page 1: ...31 F 0 to 55 C Storage 40 to 158 F 40 to 75 C Output Current sinking square wave pulse Scaled pulse output 0 1 s duration 6 1 Hz max or high alarm output or low alarm output sensor pass through pulse...

Page 2: ...Optional Protective Cover Closed PROUDLY MADE IN THE USA RTI2 Series RATE TOTAL INDICATOR INSTRUCTIONS RTI2 SERIES RATE TOTAL INDICATOR INSTRUCTIONS...

Page 3: ...eter Page 7 RTI2 139 EX Magmeter Page 7 Settings K Factor Changing Flow Indicator Settings Page 8 RTI2 Secondary Menu Functions Page 10 Troubleshooting Problems Probable Causes Things to Try Page 11 T...

Page 4: ...for the RTI2 with dual 24 and 12VDC outputs to power both the loop and sensors requiring more power than the loop can supply Pulse and 4 20mA analog outputs can be used to signal external devices e g...

Page 5: ...when loop powered Display Rate 5 digit autorange Total 8 digit Units Rate Units Gallons Second Minute Hour Day Liter Second Minute Hour Day Cubic Feet Second Minute Hour Day Cubic Meters Second Minute...

Page 6: ...nverted from a wall mount to a meter mount using the mounting kit contact distributor that includes a lower housing and associated hardware and installs as follows 1 Remove the strain relief through w...

Page 7: ..._ _ S PULSE OUT 2 PULSE OUT 1 _ _ FT440 GREEN WHITE BLACK _ FT440 139 Dual Power Supply Lower Housing 4 20mA Device _ Dashed line shows terminal connection if 4 20mA device is not used To 115Vac Power...

Page 8: ...0Vdc 12 24Vdc 12 Vdc Power Supply _ _ _ _ S White Green Red Black Red 4 20mA Device _ 9 30 Vdc Power Supply _ Red Black _ S _ RTI2 Zero Adj Status LED 1 Option Only 2 3 4 5 6 Reverse Output Forward Ou...

Page 9: ...ALLON OF FLOW All menu screens consist of two rows of tabs surrounding a dialog box that lets you view and change setup parameters Set K R UNIT SET K T UNIT SET D SET P RESET EXIT 086238 235 PRESS TO...

Page 10: ...total flow volume to zero This tab is not available when the 64 non resettable total option is ordered Exit R UNIT SET K T UNIT SET D SET P RESET EXIT PRESS TO EXIT MENU AND RETURN TO FLOW DISPLAY Re...

Page 11: ...for both outputs Either output can be changed to alarm high or alarm low If alarm options are selected menu tabs for setting the alarms will be displayed on the main menu alarm 1 or the secondary menu...

Page 12: ...or incorrect 4 20mA output RTI2 only Wrong 4mA setting or wrong 20mA setting Use Set 4 to correct target minimum flow rate Use Set 20 to correct target top flow rate Inadequate loop power supply volt...

Page 13: ...LT 14301r1 1 20160301 3 1 16 Dwyer Instruments Inc 102 Indiana Highway 212 Michigan City IN 46360 USA P 219 879 8868 F 219 872 9057 1 800 872 9141 www dwyer inst com...

Reviews: