Page 9
RTI2 INSTRUCTIONS
Dwyer Instruments, Inc. • 219.879.8868 • www.dwyer-inst.com
T Unit
R UNIT
SET K
T UNIT
SET D
SET P
RESET
EXIT
TOTAL = GALLONS
PRESS TO SET UNITS
FOR DISPLAY
View or change the total volume units
Set D
R UNIT
SET K
T UNIT
SET D
SET P
RESET
EXIT
000
PRESS TO CHOOSE
NUMBER OF DECIMAL PLACES
IN TOTAL DISPLAY
View or change the number of decimals displayed in the
total volume display
Set P or Set A
R UNIT
SET K
T UNIT
SET D
SET P
RESET
EXIT
00000.0 GALLONS
PRESS TO SET NUMBER OF
GALLONS TOTALIZED PER
PULSE SENT OUT PULSE1
The factory setting will show Set P which allows one to view
or change the volume of flow totalized per pulse sent to
pulse out 1. The units for Set P follow the units selected for
the rate display. Secondary menu selection will change the
display to Set A. The alarm can be set to trigger on either a
high or low flow condition as determined by the user.
Set 20 (RTI2 ONLY)
R UNIT
SET K
T UNIT
SET D
SET P
RESET
EXIT
00000.0 GALLONS
PRESS TO SET THE FLOW
RATE AT WHICH 20 mA
(MAX) OUTPUT IS DESIRED
SET 20
Input the flow rate at which 20 mA (max) output is desired
Reset
R UNIT
SET K
T UNIT
SET D
SET P
RESET
EXIT
PRESS TO RESET TOTAL
Reset the total flow volume to zero. This tab is not available
when the -64 non resettable total option is ordered
Exit
R UNIT
SET K
T UNIT
SET D
SET P
RESET
EXIT
PRESS TO EXIT MENU AND
RETURN TO FLOW DISPLAY
Return to the home screen, enter a submenu, or accept a
parameter change
Secondary Menu.
OUTP
MAIN
INP
P CODE
EXIT
SCROLL TO SELECT A
SUBMENU AND CHANGE
FEATURES
To enter the secondary menu use the left/right arrow keys to
navigate to the exit tab. While the exit tab is highlighted press
the up arrow 4 times. The secondary menu, shown above, will
now be displayed.
OutP
OUTP
MAIN
INP
P CODE
EXIT
PRESS TO SET ALARMS,
PULSE OUT, AND 4-20 mA
View or change the function of Set P tab on the main menu.
Options are pulse out 1, alarm low, alarm high. If the alarm
high/low option is selected a Set H (hysteresis) tab is
available. The hysteresis entry is a % value. The value defines
the % change required for a change in alarm state to occur.
InP
OUTP
MAIN
INP
P CODE
EXIT
PRESS TO SET F, J, REED
AND PULSE INDICATOR
View or change the filter (set F), jitter (set J), enable
reed mode. Use the filter setting if the display is jumping
excessively due to flow conditions. Use the jitter setting to
enter a time delay to handle start up conditions. Jitter units
are seconds.
Pcode
OUTP
MAIN
INP
P CODE
EXIT
PRESS TO SET PASSCODE
AND CHANGE PROTECTED
FEATURES
Enter the passcode for access to protected features.
SETTINGS