Dwyer Instruments RTI2 Series Instructions Manual Download Page 10

Page 9

RTI2 INSTRUCTIONS

Dwyer Instruments, Inc. • 219.879.8868 • www.dwyer-inst.com 

T Unit

 

R UNIT

SET K

T UNIT

SET D

SET P

RESET

EXIT

TOTAL = GALLONS

PRESS    TO SET UNITS

FOR DISPLAY 

View or change the total volume units

Set D

 

R UNIT

SET K

T UNIT

SET D

SET P

RESET

EXIT

000

PRESS    TO CHOOSE

NUMBER OF DECIMAL PLACES

IN TOTAL DISPLAY

View or change the number of decimals displayed in the 

total volume display

Set P or Set A

R UNIT

SET K

T UNIT

SET D

SET P

RESET

EXIT

00000.0 GALLONS

PRESS    TO SET NUMBER OF

GALLONS TOTALIZED PER

PULSE SENT OUT PULSE1

The factory setting will show Set P which allows one to view 

or change the volume of flow totalized per pulse sent to 

pulse out 1. The units for Set P follow the units selected for 

the rate display. Secondary menu selection will change the 

display to Set A. The alarm can be set to trigger on either a 

high or low flow condition as determined by the user.

Set 20   (RTI2 ONLY)

R UNIT

SET K

T UNIT

SET D

SET P

RESET

EXIT

00000.0 GALLONS

PRESS    TO SET THE FLOW 

RATE AT WHICH 20  mA

(MAX) OUTPUT IS DESIRED

SET 20

Input the flow rate at which 20 mA (max) output is desired

Reset

 

R UNIT

SET K

T UNIT

SET D

SET P

RESET

EXIT

PRESS    TO RESET TOTAL

Reset the total flow volume to zero. This tab is not available 

when the -64 non resettable total option is ordered

Exit

R UNIT

SET K

T UNIT

SET D

SET P

RESET

EXIT

PRESS    TO EXIT MENU AND 

RETURN TO FLOW DISPLAY

Return to the home screen, enter a submenu, or accept a 

parameter change

Secondary Menu.

 

OUTP

MAIN

INP

P CODE

EXIT

SCROLL TO SELECT A 

SUBMENU AND CHANGE

FEATURES

To enter the secondary menu use the left/right arrow keys to 

navigate to the exit tab. While the exit tab is highlighted press 

the up arrow 4 times. The secondary menu, shown above, will 

now be displayed. 

OutP 

 

OUTP

MAIN

INP

P CODE

EXIT

PRESS    TO SET ALARMS,

PULSE OUT, AND 4-20  mA

View or change the function of Set P tab on the main menu. 

Options are pulse out 1, alarm low, alarm high. If the alarm 

high/low option is selected a Set H (hysteresis) tab is 

available. The hysteresis entry is a % value. The value defines 

the % change required  for a change in alarm state to occur.

InP

 

OUTP

MAIN

INP

P CODE

EXIT

PRESS    TO SET F, J, REED

AND PULSE INDICATOR

View or change the filter (set F), jitter (set J), enable 

reed mode. Use the filter setting if the display is jumping 

excessively due to flow conditions. Use the jitter setting to 

enter a time delay to handle start up conditions. Jitter units 

are seconds.

Pcode

 

OUTP

MAIN

INP

P CODE

EXIT

PRESS    TO SET PASSCODE 

AND CHANGE PROTECTED 

FEATURES

Enter the passcode for access to protected features.

SETTINGS

Summary of Contents for RTI2 Series

Page 1: ...31 F 0 to 55 C Storage 40 to 158 F 40 to 75 C Output Current sinking square wave pulse Scaled pulse output 0 1 s duration 6 1 Hz max or high alarm output or low alarm output sensor pass through pulse...

Page 2: ...Optional Protective Cover Closed PROUDLY MADE IN THE USA RTI2 Series RATE TOTAL INDICATOR INSTRUCTIONS RTI2 SERIES RATE TOTAL INDICATOR INSTRUCTIONS...

Page 3: ...eter Page 7 RTI2 139 EX Magmeter Page 7 Settings K Factor Changing Flow Indicator Settings Page 8 RTI2 Secondary Menu Functions Page 10 Troubleshooting Problems Probable Causes Things to Try Page 11 T...

Page 4: ...for the RTI2 with dual 24 and 12VDC outputs to power both the loop and sensors requiring more power than the loop can supply Pulse and 4 20mA analog outputs can be used to signal external devices e g...

Page 5: ...when loop powered Display Rate 5 digit autorange Total 8 digit Units Rate Units Gallons Second Minute Hour Day Liter Second Minute Hour Day Cubic Feet Second Minute Hour Day Cubic Meters Second Minute...

Page 6: ...nverted from a wall mount to a meter mount using the mounting kit contact distributor that includes a lower housing and associated hardware and installs as follows 1 Remove the strain relief through w...

Page 7: ..._ _ S PULSE OUT 2 PULSE OUT 1 _ _ FT440 GREEN WHITE BLACK _ FT440 139 Dual Power Supply Lower Housing 4 20mA Device _ Dashed line shows terminal connection if 4 20mA device is not used To 115Vac Power...

Page 8: ...0Vdc 12 24Vdc 12 Vdc Power Supply _ _ _ _ S White Green Red Black Red 4 20mA Device _ 9 30 Vdc Power Supply _ Red Black _ S _ RTI2 Zero Adj Status LED 1 Option Only 2 3 4 5 6 Reverse Output Forward Ou...

Page 9: ...ALLON OF FLOW All menu screens consist of two rows of tabs surrounding a dialog box that lets you view and change setup parameters Set K R UNIT SET K T UNIT SET D SET P RESET EXIT 086238 235 PRESS TO...

Page 10: ...total flow volume to zero This tab is not available when the 64 non resettable total option is ordered Exit R UNIT SET K T UNIT SET D SET P RESET EXIT PRESS TO EXIT MENU AND RETURN TO FLOW DISPLAY Re...

Page 11: ...for both outputs Either output can be changed to alarm high or alarm low If alarm options are selected menu tabs for setting the alarms will be displayed on the main menu alarm 1 or the secondary menu...

Page 12: ...or incorrect 4 20mA output RTI2 only Wrong 4mA setting or wrong 20mA setting Use Set 4 to correct target minimum flow rate Use Set 20 to correct target top flow rate Inadequate loop power supply volt...

Page 13: ...LT 14301r1 1 20160301 3 1 16 Dwyer Instruments Inc 102 Indiana Highway 212 Michigan City IN 46360 USA P 219 879 8868 F 219 872 9057 1 800 872 9141 www dwyer inst com...