![DWK WHEATON R2P 2.0 Instruction Manual Download Page 45](http://html.mh-extra.com/html/dwk/wheaton-r2p-2-0/wheaton-r2p-2-0_instruction-manual_2544457045.webp)
WHEATON® R2P
™
2.0 Bottom Drive Roller Apparatus
45
DWK Life Sciences | 856.825.1100 |
www.dwk.com
| P/N 50099817 Rev. 11/2021
5
ADVANCED OPERATION
Although the R2P 2.0 network capabilities allow full monitoring and control of the unit from a remote
location, a performance history can still be extracted from the unit. A FIFO (first in first out) internal
event file can be downloaded simply using a USB flash drive. The event file is a comma delimited
ASCII file that records, with a date and time stamp, any event that changes the operation of the unit.
User changes such as bottle speed adjustments, ramp rates, bottle diameter settings, and reverse
times will be recorded. In addition, all alarm conditions will be reported at the time of initial fault, and
initial clear. The log file is capable of storing 14,336 events. The event log is a circular buffer so when
it fills, older entries will be overwritten. Due to sector sizes of the flash memory where the log
records are stored, 1,024 of the oldest records will be lost when wrap-around occurs.
FIGURE 30
– USB FLASH DRIVE PORT
USB Flash
Drive Port