background image

INTRODUCTION 

 

Th

e Water Probe is an accessory for the DURRIDGE RAD7 Electronic Radon Detector. It is a device to bring the 

radon concentration in a closed air loop into equilibrium with the radon concentration in the body of water in 
which it is submerged. It consists of a length of  hydrophobic yet gas-permeable membrane tubing, called an 
“exchanger”, that brings the radon in a closed air loop into equilibrium with the surrounding water. 

Th

e radon in 

the air loop is monitored continuously by the RAD7. 

Th

e partition coe

cient, the ratio of radon concentration in the water to that in the air at equilibrium, is 

determined by the temperature at the air/water interface. 

Th

is temperature is measured with a temperature probe 

inserted into the water. As an example, at a water temperature of 10 degrees Celsius, the partition coe

cient is 

about 1:3. 

Th

at means there is three times higher concentration of radon in the air than in the water, so there is, 

in e

ect, a gain of three times in the sensitivity of the system to radon in water, compared to radon in air. 

It takes time for the water to deliver radon to the air loop and for the RAD7 to respond to the changed radon 
concentration. With optimum con

guration the response time of the system is around 2 hours. 

Fig. 1  Assembled Water Probe 

Introduction

4

CAUTION 

There is a potential for biofouling to occur in long-term monitoring applications. The 
Water Probe should therefore be inspected periodically to ensure that there is good 
contact between the radon exchange tubing and the surrounding water.

Summary of Contents for RAD7

Page 1: ...WATER PROBE Passive Slow Response Radon in Water Accessory for the RAD7 User Manual...

Page 2: ...Protocol 6 1 4 3 User Protocol 6 2 MEASUREMENT PROCEDURE 7 2 1 Start Up 7 2 1 1 Temperature Probe 7 2 1 2 Start Measurement 7 2 2 Speed of Response 7 2 2 1 Measurement in Progress 7 2 2 2 In fl uenci...

Page 3: ...ff ect on Response Time 11 4 4 Custom DRYSTIK Settings 11 Fig 4 Water Probe con fi guration with Active DRYSTIK 12 5 CARE MAINTENANCE AND TROUBLESHOOTING 13 5 1 Water Catastrophe 13 5 2 RAD7 Care 13...

Page 4: ...mperature is measured with a temperature probe inserted into the water As an example at a water temperature of 10 degrees Celsius the partition coe ffi cient is about 1 3 Th at means there is three ti...

Page 5: ...t this end to the air inlet fi lter with 1 8 ID tubing at the fi lter end which is then placed on the RAD7 INLET Th e Luer taper ensures an airtight connection Fig 2 RAD7 Water Probe Standard Setup 1...

Page 6: ...tteries in a fully charged condition Set the RAD7 s Mode to Auto which starts the instrument in Sni ff mode counting 218 polonium decays only and switched automatically to Normal Mode counting 218 pol...

Page 7: ...a standard deviation of 10 At very low concentrations it may take hours and averaging over many cycles to reach a su ffi ciently precise value 2 2 2 In fl uencing Factors Th ere are two processes req...

Page 8: ...the process and which must be replenished Th e most obvious is the desiccant A new or regenerated Laboratory Drying Unit will normally last for about ten days of continuous use in a temperate climate...

Page 9: ...ce as instructed Th e water temperature information is used by CAPTURE to calculate the radon in water concentrations For detailed instructions please see the Water Probe information in the CAPTURE us...

Page 10: ...formula is applied automatically in CAPTURE when the Water Type is set to Fresh Water in the Run Parameters Window as shown in Fig 3 below 3 2 2 Schubert Et Al Formula If the Water Type is set to Sal...

Page 11: ...It should be inserted between the DRYSTIK air outlet and the inlet fi lter on the RAD7 as shown in Fig 4 on the following page Th e desiccant will last for a very long time and the drying tube will ad...

Page 12: ...Fig 4 Water Probe con fi guration with Active DRYSTIK Section 4 DRYSTIK 12...

Page 13: ...id onto its hinges Th e wrap will make the RAD7 almost watertight If it tears it can be easily replaced at any time Th e instrument should in any case be returned every year for recalibration It is us...

Page 14: ...dioxide measurements in coastal waters Environmental Science and Technology 46 14 7685 7691 Schubert M Paschke A Lieberman E Burnett W C 2012 Air Water Partitioning of 222Rn and its Dependence on Wat...

Page 15: ...978 667 9556 Fax 978 667 9557 Web www durridge com Email service durridge com Copyright 2021 DURRIDGE Company Inc All rights reserved DURRIDGE the DURRIDGE logo and the Works with the DURRIDGE logo ar...

Reviews: