![Durr Dental V 300 S Installation And Operating Instructions Manual Download Page 26](http://html.mh-extra.com/html/durr-dental/v-300-s/v-300-s_installation-and-operating-instructions-manual_2542911026.webp)
5.2
VS 300 S
1
2
3
4
5
6
7
8
9
1
Motor
2
Auxiliary air nozzle
3
Exhaust air muffler
4
Control electronics
5
Waste water connection
6
Separation
7
Suction connection
8
Turbine wheel
9
Exhaust air connection
The VS suction units are used in wet suction systems. The suction units can be installed on the same
floor as the treatment units or on the floor underneath. The necessary air flow and vacuum are gener-
ated by a rapidly rotating impeller. An auxiliary air nozzle on the turbine housing protects the suction
units against overheating.
When an appropriate vacuum for the machine is applied, approx. 300 l/min of air is sucked in through
the suction cannula.
The impeller, the separation turbine and the waste water pump are driven by the motor.
The mixture of liquids, solid particles and air drawn in passes through the inlet connection and into the
suction unit. The coarse filter holds back the solid particles.
Product description
24
7119100007L02 1909V001
EN