![Durr Dental V 300 S Installation And Operating Instructions Manual Download Page 15](http://html.mh-extra.com/html/durr-dental/v-300-s/v-300-s_installation-and-operating-instructions-manual_2542911015.webp)
4.2
VS 300 S
Electrical data
7122-01
7122-04
7122-02
7122-03
Rated voltage
V
230, 1~
230, 1~
230, 1~
Mains frequency
Hz
50
50
60
60
Nominal current
A
2.9
2.9
3.7
3.7
Starting current, approx.
A
10.4
10.4
9.5
9.5
Motor protection
Motor winding overheat protector 160 °C
(±5 °C)
Rated power
W
580
580
800
800
Type of protection
IP 24
Protection class
I
Protective low voltage
V
24 ~
Output
VA
4
Connections
Suction connection, DürrConnect special
mm
Æ
30
Exhaust air connection (external)
mm
Æ
30
Drain connection, DürrConnect
mm
Æ
20
Media
Max. unimpeded flow rate
l/min
700
700
800
800
Max. suction system pressure *
mbar/hPa
-200
Max. rate of flow of fluids
l/min
4
Max. suction height
cm
50
*
Depending on unit type
General data
Duty cycle
%
100
Dimensions (H x W x D) *
cm
38 x 31 x 32
Weight, approx.
without housing
with housing
kg
kg
13.5
21.5
Noise level ** approx.
without housing
with housing
dB(A)
dB(A)
63
51
63
51
65
54
65
54
*
Values without accessories and add-on parts
**
Noise level in accordance with ISO 3746
Ambient conditions during storage and transport
Temperature
°C
-10 to +60
Product description
7119100007L02 1909V001
13