After the end of the rinsing process drain any
remaining fluid in the system.
Disinfect using Vector/RinsEndo Disinfection:
❯
Pour 2 sealing caps (approx. 40 ml) of
Vec-
tor/RinsEndo disinfection
into the fluid con-
tainer.
2x
❯
Touch the RINSE (rinsing/disinfecting) button
for at least 2 seconds.
RINSE
❯
LED flashes – the
Vector/RinsEndo Disinfec-
tion
is rinsed into the system and the proc-
ess ends automatically. The
Vector/
RinsEndo Disinfection
remains in the system
until the next treatment.
❯
Empty any remaining
Vector/RinsEndo Dis-
infection
from the fluid container.
❯
Thoroughly rinse the fluid container with
water and dry.
14.2 Initial start-up after a break in
treatment for more than 24
hours
The initial start-up process will depend upon
whether the hose system was reprocessed
before the break in treatment. Proceed as follows
depending on the situation:
1. Reprocessing performed before the treat-
ment break:
❯
Rinse the system with water.
Thorough rinsing with water will flush out
any disinfectant remaining in the hoses,
and any irritation to the patient caused by
the taste of residual disinfectant will be
avoided.
❯
Pull off the hose connection from the hand-
piece.
❯
Lay the handpiece hose in the sink.
❯
Fill the fluid container with water.
❯
Touch the RINSE (rinsing/disinfecting) button
for at least 2 seconds.
LED flashes – the rinsing process takes
around 30 seconds and ends automatically.
2. Decontamination process was not carried
out before surgery break:
❯
Perform reprocessing before initial start-up of
the unit, "13 Reprocessing".
Usage
9000-615-32/02 2012V003
43