12.2 Activating the cleaning proc-
ess of the device
We recommend that cleaning is per-
formed every 4 weeks. The cleaning
process can be started at any time as
required.
After approx. 30 operating the CLEAN LED will
light up on the operating panel to show that
cleaning is required.
A cleaning cycle comprises two steps that need
to be started one after the other:
ü
CLEAN
ü
RINSE (rinsing/disinfecting)
The cleaning process is not finished until
both steps have been successfully per-
formed one after the other.
CLEAN:
❯
Pull off the hose connection from the hand-
piece.
❯
Place the handpiece hose in a sink or in a
suitable container.
❯
Pour 2 sealing caps (approx. 40 ml) of
Vector
cleaner
undiluted into the empty fluid con-
tainer.
2x
❯
Touch the CLEAN button for at least 2 sec-
onds.
CLEAN
1x
+
+
LED CLEAN and the LED in the fluid container
flash and a clicking sound is heard.
The device is cleaned for approx. 10 minutes
with
Vector cleaner
until the fluid container is
empty. The process ends automatically.
LED RINSE lights up and a cyclically repeating
warning signal sounds as an indication that the
device must be rinsed with water after the clean-
ing with
Vector cleaner
.
+
...
RINSE
45 s
45 s
3x
3x
LED CLEAN and the LED in the fluid container
flash and a clicking sound is heard.
The device is cleaned for approx. 10 minutes
with
Vector cleaner
until the fluid container is
empty. The process ends automatically.
LED RINSE lights up and a cyclically repeating
warning signal sounds as an indication that the
device must be rinsed with water after the clean-
ing with
Vector cleaner
.
Usage