10
11
Immunity Test Standard
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, interruptions and
variations on power supply input
lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
IEC 60601-1-2 Test Level
+8kV contact
+2kV, +4kV, +8kV, +15kV air
+2kV, for power supply lines
+1kV signal input/ output
100 kHz repetition frequency
+0.5kV, +1kV differential lines
+0.5kV, +1kV, +1kV common
mode
0 % UT; 0.5 cycle. At 0°, 45°, 90°,
135°, 180°, 225°, 270° and 315°.
0 % UT; 1 cycle and 70 % UT;
25/30 cycles; Single phase: at 0°.
0 % UT; 250/300 cycle
30 A/m
50Hz/ 60Hz
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio
bands between 0,15 MHz and 80
MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz 80 % AM at 1
kHz
Compliance Level
+8 kV contact
+2kV, +4kV, +8kV, +15kV air
+2kV, for power supply lines
Not applicable
100 kHz repetition frequency
+0.5kV, +1kV differential lines
Not Applicable
0 % UT; 0.5 cycle. At 0°, 45°,
90°, 135°, 180°, 225°, 270°
and 315°.
0 % UT; 1 cycle and 70 % UT;
25/30 cycles;
Single phase: at 0°.
0 % UT; 250/300 cycle
30 A/m
50Hz/ 60Hz
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio
bands between 0,15 MHz and
80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz 80 % AM
at 1 kHz
Note: UT is the a.c. mians voltage prior to application of the test level.
Warning instruction indicating a
risk of injury or damage to health
Read operating
instructions before use
Caution: Consult
accompanying documents
Class II Equipment
Date of Manufacture
Manufacturer
Keep Dry
Safety note indicating possible
damage to the appliance/ acces-
Specifies serial number of the
“Daylight”
OFF/ ON
Authorized Representative in the
European Community
Ingress Protection Rating
O / I
IP20
EC/REP
The following symbols appear in this manual, on the “Daylight” or on its accessories:
1. Screen
2. Brackets
3. Mounting hole
4. Carrying handle
5. Fold-out stand
6. On/ Off Switch
7. Retaining Clips
8. Reflector
9. Tubes, 2 x 36W
Unit Description
Setup
1. Remove the appliance from the packaging.
2. Fold the stand out and place the appliance on a solid surface, such as a
table.
3. Place the appliance approximately 20-60 cm from your eyes, the light must
bathe your face from the side.
Mains Connection
Connect the appliance only to the mains voltage listed on the type plate.
Push the plug completely into the mains socket.
Note
•
Push the plug completely into the mains socket.
•
Ensure that there is a socket near the place where the appliance will
stand.
•
Lay the mains cable in such a way that no one can trip over it.
Fastening of Daylight on the wall
There is a mounting hole on the back side of the appliance. At this mounting
hole the appliance can be fixed on the wall with screw. Depending on the
type of wall, an appropriate screw must be used, may be with pegs. E.g. for
concrete wall, M4, 25mm long Stainless Steel Self-Tapping Screw which is
screwed into a anchor.
The screw and the anchor are not included in the scope of delivery.