Fault Code
Description
Action Required
16
Front panel high ambient
temperature alarm
Alarm –
Ambient temperature at front panel is higher than high
ambient temperature limit. Occurs when ambient temperature
exceeds the set ambient limit by 5°C or more. Compressor, heater,
fan, and pump are turned off.
Lower temperature in area where Chiller is located or increase high
ambient temperature limit value.
Enabling / Disabling the Local Lockout
This feature is used to prevent unauthorized or accidental changes to set point and other operational values.
When enabled, the values for the following functions can be displayed, but not changed:
•
Temperature unit
•
Temperature set point
•
Pressure / flow rate units
To enable the local lockout, press and hold the Select/Set Knob until LLO is displayed (approximately 5 seconds).
Once enabled, LLO will appear momentarily when the Select/Set Knob is pressed to display the set point.
To disable the local lockout, press and hold the Select/Set Knob until CAn appears momentarily as local lockout
status changes from enabled (LLO) to disabled (approximately 5 seconds).
IMPORTANT:
The Local Lockout feature does not prevent set point changes entered via the RS232 interface.
Summary of Contents for DCA203C4
Page 1: ...Operator s Manual DuraChill Air and Water Cooled 2 and 3 HP Chillers 110 275 25 February 2015...
Page 9: ...110 275 8 Rear View Power Switch Power Switch...
Page 36: ...110 275 35 Pump Performance Specifications subject to change without notice...
Page 40: ...110 275 39 Appendix Flow Diagram Air Cooled Chillers...
Page 41: ...110 275 40 Flow Diagram Water Cooled Chillers...
Page 42: ...110 275 41 Wiring Diagram 208 230V 1 Phase Air Cooled Chillers...
Page 43: ...110 275 42 Wiring Diagram 208 230V 380 460V 3 Phase Air and Water Cooled Chillers...