Footswitch Operation

The one-touch functions such as [

EMPTY

] and [

FILL

] cannot be done with the

footswitch. The same phenomenon will occur if you continue to push the pedals for

functions such as [

EMPTY

] and [

FILL

]. If you ease the pressure off the pedal, the

motor will stop.

MANUAL 

]

Mode

In this mode, the micropipette is manually

filled or emptied.

Touch the [

MANUAL

] icon to go to the

MANUAL MODE screen in the diagram

below.

EMPTY Function

When you touch this icon and the icon

changes to a red [

STOP

] icon, the motor

continues to empty the micropipette.

When the plunger has extended completely

(approximately 23 mm from the end of the

collet), there will be a beeping sound and

the screen will return to the [

EMPTY

] icon

at the same time.

To stop it in the middle, push [

STOP

].

Summary of Contents for Nanoject II

Page 1: ...Nanoject II Digital Control Box MADE IN USA INSTRUCTION MANUAL 3 000 029 D LABORATORY EQUIPMENT E21997...

Page 2: ...E ATTENTION TO THE RULES FOR SAFE OPERATION WARNINGS AND CAUTIONS Rules for Safe Operation Connections to Power Source Check the power source to see that the line voltage corresponds to the voltage in...

Page 3: ...le 3 000 027 Footswitch 3 000 032 remote capability for fill empty and inject functions Micromanipulator MM33 right hand 3 000 024 R left hand 3 000 024 L Support Base 3 000 025 SB Magnetic Base 3 000...

Page 4: ...filled with oil or other non compressible fluid before attachment to the injector Silicone or mineral oil is frequently used Backfilling is facilitated by using the 30 gauge x 2 needle and a syringe D...

Page 5: ...ing the FILL icon again The plunger is fully retracted to its preset home position when you hear an audible beep Viscous samples may require you to fill in small steps allowing the sample to equilibra...

Page 6: ...iple injections can be performed by simply pressing the INJECT icon again and again The injection rate is independent of the fill rate as programmed in the SETUP mode Press the EXIT icon to exit back...

Page 7: ...g the number of injection cycles Again use the and icons to select the value desired Pressing the BACK icon always returns you to the previously displayed screen Pressing the NEXT icon once again brin...

Page 8: ...ccur if plunger is not stopped at proper Home position Press the HOME icon to manually retract the plunger to the Home position Press the EXIT icon to exit back to the operational mode screen Below th...

Page 9: ...e In this mode the micropipette is manually filled or emptied Touch the MANUAL icon to go to the MANUAL MODE screen in the diagram below EMPTY Function When you touch this icon and the icon changes to...

Page 10: ...In this mode the FILL and EMPTY pedals are disabled Footswitch Operation INJECT Mode On the Operation screen select INJECT mode Enter the manual INJECT conditions VOLUME NL Displayed in ranges within...

Page 11: ...11 Notes...

Page 12: ...ev 0219 Printed in USA 500 Parkway Box 700 Broomall PA 19008 800 523 7480 FAX 610 353 6204 Website www drummondsci com For Technical Assistance call 1 800 523 7480 or e mail info drummondsci com Quest...

Reviews: