With proper care and maintenance, your centrifuge will provide years of laboratory service. For proper care, the 
following steps should be taken:

 

o

 

Always Spin Balanced Loads: 

Make certain that you are always spinning a balanced load, as shown in the 

previous section. These centrifuges have a unique counter balanced motor mounting design which produces 
excellent vibration dampening. However, out–of–balance loads may break glass test tubes and may produce 
unsatisfactory separation results. Proper load balancing will improve sample separation and extend the life of 
the centrifuge.  

o

 

Motor and Electrical Maintenance: 

The highest quality electrical components have been selected for the DASH 

Apex centrifuges and should not need maintenance or servicing for the life of the centrifuge. 

o

 

Tube Holder Replacement: 

It is recommended that the tube holders be replaced after 24 months of use. 

Inspect tube holders regularly for cracks. If cracks are discovered, replace immediately. 

o

 

Remove Accessories Before Moving: 

All tube holders, samples, and caps must be removed from the rotor 

chamber before transporting or storing the centrifuge to prevent damage and injury. 

 

C

LEANING AND 

D

ISINFECTION

 

To prolong the life of the centrifuge, cleaning and disinfection is recommended every six months or whenever there 
is a spillage or tube breakage. Contaminants must be removed immediately or corrosion and premature 
degradation of components can occur. Before using any cleaning or decontamination methods other than those 
recommended by the manufacturer, users should verify with the manufacturer that the proposed method will not 
damage the equipment.

 

o

 

Unplug the centrifuge before cleaning. 

o

 

Use appropriate personal protective equipment (PPE). 

o

 

Apply cleaning solutions with a towel or cloth. Do not submerge the centrifuge in water or other cleaning 

solutions as this will cause damage and void the warranty. 

o

 

ONLY isopropyl alcohol or a 10% (5500 PPM) bleach solution should be used to disinfect the centrifuge and its 

accessories. 

o

 

All surfaces must be dried immediately after cleaning and disinfecting. 

    

TBQ

 

G

ERMICIDAL PRODUCTS ARE NOT RECOMMENDED AS THEY MAY CAUSE DAMAGE TO THE 

CENTRIFUGE

.

 

W

IPE OFF THOROUGHLY AFTER USE TO PREVENT VOIDING THE WARRANTY

 

o

 

Fully/partially halogenated hydrocarbons, ketones,  esters, ethers, benzyls, ethyl benzenes, and all other 

chemicals not prescribed by the manufacturer shall not be used as they may cause damage to the rotor chamber, 

rotor, tube holders, accessories and centrifuge exterior and void the warranty.

 

 

 

Summary of Contents for DASH COAG

Page 1: ...DASH COAG Operator s Manual...

Page 2: ......

Page 3: ...ial Setup 2 Quick Start 3 Settings 4 Review Cycle Time and Speed Settings 4 Changing Cycle Time and Speed Settings 4 Operation 5 Balancing Loads 5 Care and Preventative Maintenance 6 Cleaning and Disi...

Page 4: ......

Page 5: ......

Page 6: ......

Page 7: ...to reinforce the tube holders and provide high strength and durability o A clear lid permits safe observation of samples and optical calibration of speed o The lid safety system only allows entry into...

Page 8: ...repairs have been performed This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with this operator manual may cause interference to radio c...

Page 9: ...ng the locking mechanism Entry is only possible when the rotor is stopped Stop Pressing the UNLOCK button during operation will terminate the run and unlock the lid after the rotor has come to astop C...

Page 10: ...in the current cycle 10 beeps flashes equal 10 minutes of run time Run time starts when the rotor reaches 90 of desired speed and stops when the rotor starts decelerating o Pressing the START button a...

Page 11: ...will slow to a complete stop and the lid light will flash o The locking mechanism will disengage for 60 seconds allowing entry into the rotor chamber To unlock after more than 60 seconds have elapsed...

Page 12: ...tingor storing the centrifuge to prevent damage andinjury CLEANING AND DISINFECTION To prolong the life of the centrifuge cleaning and disinfection is recommended every six months or whenever there is...

Page 13: ...mage o If the tube holders or buckets have any damage even slight safely dispose of them and replace them o If the rotor appears damaged contact Customer Service for further assistance Excessive noise...

Page 14: ...o change the run preset time The lid does not open o Ensure that the lid knob is turned fully counterclockwise o If the knob cannot be turned counterclockwise turn it fully clockwise press UNLOCK and...

Page 15: ...0 Watts Centrifuge Motor H P Brushless DC Max G Force 4 400 xg Max Speed 6 600 RPM Min Cycle Time 1 minute Max Cycle Time 30 minutes Use only with approved accessories from the original manufacturer A...

Page 16: ...ectronic equipment The separate collection and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protect...

Reviews: