The cycle time and speed are not set
to the desired value
o
Check the setting by following the instructions in the section on
Changing Cycle Settings. If the preset is not the desired length,
follow the procedure on the same page to change the run
preset time.
The lid does not open
o
Ensure that the lid knob is turned fully counterclockwise.
o
If the knob cannot be turned counterclockwise, turn it fully
clockwise, press UNLOCK, and turn counterclockwise.
o
If the lid remains locked after this and will not unlock, the
electronics may have been damaged. Contact customer service
for assistance.
Clicking noise during braking gets loud
o
Make sure that the screw in the center of the rotor is tight.
Lid does not stay up
o
Tighten the center screw on the lid hinge.