The LINK (Green) lamp does not light up when the detector is not
registered or the communication connection is not established.
When the READY (Green) and AP (Blue) lamp are blinking and LINK
(Green) lamp does not light up, a communication error has occurred.
3.5.2.3 Conduct examination
For details about operation, refer to the User Manual for the EConsole1.
①
Select the patient information or protocols on the
screen and start the examination.
- The READY lamp color is changed from green to orange when the detector and EConsole1
change to exposure ready status.
- Arrange the patient in the correct posture and position the detector aligning it with the target
body part.
- Position the X-ray generator to adjust the exposure field.
- Check all conditions before exposure.
- Make sure that Ready (Green) and LINK (Green) lamps are lit and AP lamp (Blue) is blinking.
This means that the system is ready to start an examination
A communication error has occurred when LINK lamp blinks.
When the READY lamp (Orange) blinks slowly, the detector is in detector
selection status (Sleep).
The detector enters detector selection status automatically when it has
not been used for a certain period of time.
②
Press the exposure switch of the X-ray generator. Images captured with the
detector are transmitted to the Ecali1and appear on the monitor.