Droper VET Operating Instructions Manual Download Page 9

9/12 

3. OPERATING ISSUES 

3.1 

Stop, insufficient or absence of administration 

 

due to a device malfunction 

If the administration stops due to a device malfunction, characterized by a failure in bag pressurization, 
this incident may be confirmed by a lack of flow of the product administered in the drip chamber of 
the line in use. 

Firstly check that the cover of the Droper Vet has been closed correctly. Where necessary, open and 
re-close the cover. 

In the event that the cover is closed correctly, open the Droper Vet, unfasten the mobile plate from its 
tightening clips, check the following points and implement the relevant actions

 

:  

 

Check that there are no objects blocking the mechanism. If this is the case, remove the object 



Reinstall the mobile plate onto the mechanism by replacing the clips using pressure. Re-

pressurize the system.

 

 

Check that one or several springs have not stopped working 



Remove the bag of fluid and continue administration by gravity or manual pressure on 

the bag, or change device. Send the Droper Vet to the maintenance department for repair. 

 

Check whether the clamp, roller clamp or flow regulator is set to the OFF position 



Reposition the clamp, roller clamp or flow regulator to the desired position.

 

 

Check whether the infusion / transfusion line is being pinched by excessive bending of the latter 



Remove this bend.

 

 

Check whether the infusion / transfusion line is being clamped by an object crushing the line 
 



Remove this object.

 

 

Check whether the infusion / transfusion line has been clamped between the cover and the 
plate when arming the Droper Vet 
 



Open the Droper Vet and move the infusion / transfusion line.

 

3.2 

Stop, insufficient or absence of administration 
due to a catheter obstruction 

After checking the correct operation of the device, check whether the administration has stopped, is 
insufficient or is absent by observing the slowing or absence of flow of the infused fluid in the drip 
chamber  



 Reinsert another catheter

Summary of Contents for VET

Page 1: ...OPERATING INSTRUCTIONS DROPER VET MAHUSACA sprl Lieu dit Maison Bois 1 4877 OLNE Belgique info droper be ...

Page 2: ...of contents 1 Description of the Droper Vet 3 2 Using the Droper Vet 4 3 Operating issues 9 4 Cleaning Disinfection 10 5 Periodic inspection 11 6 Storage 11 7 Warranty conditions 12 8 Life 12 9 Useful contacts 12 ...

Page 3: ...3 12 1 DESCRIPTION OF THE DROPER ...

Page 4: ...n devices working under pressure like the pressure cuffs Types of infusion lines Two types of lines can be accepted 1 Classic infusion lines similar to these used with the pressure cuffs These lines have normally a drip chamber directly connected to the spike see attached drawing With this accessory there are two possible scenarios a Droper used residentially with flow rate monitored through the d...

Page 5: ...this accessory there is one possible scenario a Droper used residentially or ambulatory with flow regulation through the drip chamber Possible with all flexible crystalloid bag sizes up to 1000 ml Classical air line purge the Droper can be used vertically hanged or horizontally laid down with its drip chamber directed downward see aside drawing For safety reason any infused animal transfer with a ...

Page 6: ...the cover then open the cover 2 2 Inserting the infusion line Check that the choke or flow regulator is set to the OFF position Insert the spike into the infusion bag 2 3 Placing the infusion bag connected with the infusion line and closing the device 1 Insertion on the plate With the cover fully open position the bag to be administered on the plate and attach it to one of the fastening hooks on t...

Page 7: ...ged side down in order to correctly position the cover This operation pressurizes the bag 3 Positioning On the standing horse the Droper vet will be hanged in the upper part of a surcingle see under drawings by the mean of a hook not supplied that will go through the hanging ring installed on the rear side of the Droper Vet in order to benefit from a stable positioning On lying animal ideally the ...

Page 8: ...he inverted pyramid drawn on the front surface of the mobile plate of the equipment when the final peak will be visible coupled with the absence of flow of infusion solution in the drip chamber if necessary by observance of blood reflux in the tubing with the device being located below the level on which the patient is lying Once administration is complete open the Droper Vet using the same method...

Page 9: ...ot stopped working Remove the bag of fluid and continue administration by gravity or manual pressure on the bag or change device Send the Droper Vet to the maintenance department for repair Check whether the clamp roller clamp or flow regulator is set to the OFF position Reposition the clamp roller clamp or flow regulator to the desired position Check whether the infusion transfusion line is being...

Page 10: ...ssembly and cleaning the different parts removed according to the procedure described above in point 4 1 Cleaning can be disinfected by soaking scrubbing or spraying appropriate products The use of disinfectants applied using SPRAYS is ideal for routine disinfection operations and must take place in compliance with the disinfectants manufacturer s recommendations and at a distance of 30 cm from th...

Page 11: ...o operate more than 6 000 times under normal operating conditions However the manufacturer advises that users perform annual maintenance and inspections to ensure the correct operation of the device either by sending it to the manufacturer or distributor or by using the method described below Place a 1000 ml bag of NaCl 0 9 the design of which corresponds to the plate s surface e g 1000 ml Viaflo ...

Page 12: ...d or repaired by persons not authorized by us The serial number of the device must not have been modified or erased For item returns and repairs please contact the After Sales Service 8 LIFE The Droper Vet was designed to operate for a period of 2 years calculated based on a daily usage rate of 8 infusions per day from its commissioning date and for a period of 10 years from its date of manufactur...

Reviews: