20
Electrical safety standards:
IEC 60601-1: 2005
EMC:
IEC
60601-1-2:2007
Electrical shock protection:
Class 1
Liquid ingress protection:
IPX4
The cot is only suitable for use when:
Ambient temperature:
+5°C to +40°C
Relative humidity:
20% - 90% at 30°C - not condensing
Atmospheric pressure:
70kPa to 106kPa (Rated to operate at ≤ 2000m)
Pollution:
Degree 2
The cot is compliant to the requirement of 60601-1-2: 2007, the RF emissions are
very low and are not likely to cause any interference to nearby electronic equipment,
however interference to sensitive equipment is possible.
Qualified users of the cot can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment
(transmitters). If the cot or any alternative equipment is found to be operating
abnormally mitigation measures are to be taken, such as the separation distances being
increased and/or the products being re-orientated.