background image

3362443 (A3-D-P) – Page 3 of 4

PSS 5000 Series

Self-contained breathing apparatus with Sentinel TX Gauge

Instructions for Use

i

Symptom

Fault

Remedy

Face mask air leak

Lung demand regulator O-ring leaking

Renew or lubricate O-ring

Head straps not tight

Tighten

Exhalation valve leaking

Service task

Speech diaphragm defective

Service task

Unsatisfactory communication

Speech diaphragm defective

Service task

High-pressure air leak or failed leak test

Loose or dirty connector

Disconnect, clean and reconnect couplings and 

retest

Faulty hose or component

Substitute user replaceable accessories and 

retest

Air leak from medium-pressure hose connection 

at the first-stage regulator 

(safety relief valve)

Faulty O-ring, retainer, spring or first-stage 

regulator

Service task

Air leak from lung demand regulator

Ice particles on sealing elements

Press the front button (Fig 3, Item 2), allow a 

rush of air to pass through the regulator, then 

quickly press the reset button (Fig 3, Item 1) to 

switch off the positive pressure.

Air leak from quick connect cylinder coupling

Ice particles on sealing elements

Disconnect then reconnect the cylinder to the 

breathing apparatus (see the Quick Connect 

Cylinder Coupling Instructions for Use) and 

retest.

Lung demand regulator allowing constant air 

flow into the face mask

Bypass button engaged

Turn off the bypass button (Fig 3, Item 3)

Internal fault

Service task

High or low medium pressure

First-stage regulator fault

Service task

Poor sounding whistle

Whistle dirty

Clean whistle flute and retest

Whistle not functioning correctly

Activation mechanism fault

Service task

TX Gauge fails to switch on

Low batteries

Replace the batteries (see Section 3.5.3)

Unit unserviceable

Service task

Flashing amber LED every five seconds

Low batteries

Replace the batteries 

(see Section 3.5.3)

Flashing of the amber LED every 2 seconds for 

10 seconds

Failed self check, or hardware fault

Service task

TX Gauge face difficult to see through window

Dirt on gauge face window

Clean gauge face window (see Section 5.2)

Gauge face window damaged

Service task

Maintenance

5.1  Maintenance table

Service and test the breathing apparatus, including out-of-use apparatus, in accordance with the maintenance table. Record all service details and 

testing. Refer also to the Instructions for Use for the lung demand regulator, face mask and other associated equipment.

Additional inspection and testing may be required in the country of use to ensure compliance with national regulations.

Component/ System Task

After use

Every 

month

Every year

Complete apparatus

Clean and disinfect (see Section 5.2)

Visual inspection (see Section 3.5.1)

Functional testing (see Section 3.6)

Flow and static tests (see Note 1)

Lung demand 

regulator

Clean and disinfect (see Note 3 and Section 5.2)

First-stage regulator

Medium-pressure check (see Note 1)

Inspect the sintered filter (see Note 1 and Note 2)

Inspect the high-pressure connector O-ring (see Note 1 and Note 4)

Cylinder

Charge cylinder to correct working pressure

Check charged pressure (stored cylinders only)

Check test date of cylinder (carbon composite cylinders over 15 years old must be 

retired)

Recertification

According to national regulations in the 

country of use

Cylinder valve

Overhaul

At the time of cylinder recertification

Notes

○  Dräger recommendations

1.  These maintenance tasks may only be carried out by Dräger or trained service personnel. Details of the tests are contained in the Technical Manual 

which is issued to service personnel that have attended a relevant Dräger maintenance course.

2.  Replace the sintered filter if a drop in first-stage regulator performance is observed during a flow check or if it is visibly damaged.

3.  Lightly lubricate the O-ring of the lung demand regulator as required (recommended lubricant is Dow Corning 111 Valve Lubricant and Sealant). 

Products other than the recommended lubricant are not tested and may damage the equipment.

4.  Replace the high-pressure connector O-ring if it is found to leak during functional testing or if the O-ring is visibly damaged.

5.2  Cleaning and disinfecting

!

CAUTION

Cleaning agents and disinfectants listed below are not manufactured by Dräger and have been reviewed only for compatibility when used to 

clean or disinfect the subject Dräger product(s). Read and comply with all instructions for use provided by the manufacturers of such agents 

and disinfectants. Dräger expressly disclaims all responsibility for any damage, personal injury or loss resulting from the use of such agents or 

disinfectants.

Do not exceed 86 °F (30 °C) for washing, disinfecting and rinsing solutions. Do not exceed 140 °F (60 °C) for drying, and remove components 

from the drying facility immediately when dry. Drying time in a heated dryer must not exceed 30 minutes.

Do not immerse pneumatic or electronic components in cleaning solutions or water.

If water is trapped and then freezes inside the pneumatic system of the breathing apparatus (such as the lung demand regulator) its operation will 

be impaired. Prevent any liquid from entering, and thoroughly dry the breathing apparatus after cleaning to prevent this from occurring.

Refer also to the Instructions for Use for the lung demand regulator, face mask and other associated equipment.

5.2.1  Manual cleaning of the breathing apparatus (USA)

Cleaning and disinfecting materials:

 

● Cleaning agent – 1008 Green Liquid Hand Dish Wash

 

● Disinfecting  agent  –  800  Spur-Tex  Disinfectant  Cleaner-Deodorant 

(concentration: 1.6 % (2 fl oz per gallon))

 

● Use only clean lint-free cloths.

1.  Prepare cleaning solution as per manufacturer’s instructions. Clean 

the breathing apparatus manually using a cloth moistened with 

cleaning solution to remove excess dirt.

2.  Prepare disinfecting solution as per manufacturer’s instructions. Apply 

to all internal and external surfaces, ensuring that all surfaces remain 

visibly wet for 15 minutes.

3.  Rinse all components thoroughly with clean water to remove all 

cleaning and disinfecting agents.

4.  Dry all components using a dry cloth, in a heated dryer or in air.

5.  Contact service personnel or Dräger if disassembly of pneumatic or 

electronic components is required.

5.2.2  Manual cleaning of the breathing apparatus (Canada)

Cleaning and disinfecting materials:

 

● Cleaning agent – mild soap solution

 

● Disinfecting agent – Neutral Disinfectant Cleaner (concentration: 0.5 % 

(0.5 fl oz per gallon or 15 ml per 3.785 liters))

 

● Use only clean lint-free cloths.

1.  Prepare cleaning solution as per manufacturer’s instructions. Clean 

the breathing apparatus manually using a cloth moistened with 

cleaning solution to remove excess dirt.

2.  Prepare disinfecting solution as per manufacturer’s instructions. Apply 

to all internal and external surfaces, ensuring that all surfaces remain 

visibly wet for 10 minutes.

3.  Rinse all components thoroughly with clean water to remove all 

cleaning and disinfecting agents.

4.  Dry all components using a dry cloth, in a heated dryer or in air.

5.  Contact service personnel or Dräger if disassembly of pneumatic or 

electronic components is required.

5.2.3  Carrying harness – Thorough cleaning

!

CAUTION

Thorough cleaning may only be carried out by suitably trained 

personnel. Disassembly of the breathing apparatus by untrained 

personnel is not permitted and may damage the apparatus.

1.  Remove the harness from the carrying system and clean using one of 

the following methods:

a.  Clean manually in a bath containing recommended cleaning or 

disinfecting agents.

b.  Machine wash (at 86 °F (30 °C)) using a proprietary brand washing 

solution (do not use biological washing powder).

2.  Rinse the harness thoroughly with clean water to remove all cleaning 

and disinfecting agents.

3.  Dry all components including internal parts.

5.3  Maintenance work

5.3.1  Air cylinder charging

!

WARNING

Air quality for compressed-air cylinders must conform to the 

minimum grade requirements for Type 1 gaseous air as defined 

in  the  CGA  Commodity  Specification  for Air,  G-7.1  (Grade  D  or 

higher quality) and, where appropriate, be in accordance with: 

NFPA 1989 Standard on Breathing Air Quality for Emergency 

Services Respiratory Protection.

Refer to the instructions supplied with the cylinder and the charging 

apparatus for recharging a compressed-air cylinder.

Storage

6.1  Storage preparation

 

● Extend the shoulder straps, waist belt and the straps of the face mask.

 

● For storage, place the face mask in a protective bag (contact Dräger 

for supply of a suitable bag).

 

● Route rubber hoses in such a way that the bend radius is not too acute 

and the hose is not stretched, compressed or twisted.

 

● With the system switched off, a small amount of battery power is 

consumed. If the system is not to be used for a long period, remove 

the batteries (see Section 3.5.3).

6.2  Storage conditions

 

● Store the equipment between 5 and 77 °F (-15 and +25 °C). Ensure 

that the environment is dry, free from dust and dirt, and does not 

subject the equipment to wear or damage due to abrasion. Do not 

store the equipment in direct sunlight.

 

● Fix the breathing apparatus securely to any raised mounting point to 

prevent it from falling.

Disposal

Dispose of used batteries in accordance with national or local regulations. 

When required, dispose of other parts of the breathing apparatus, 

including electrical and electronic equipment, in line with any national or 

local environmental regulations.

Technical data

Compressed-air cylinders:

 

● 30 minutes to 60 minutes capacity

 

● 2216 psi or 4500 psi pressure

 

● Composite materials

Operating pressures:

 

● Gauge operating pressure: 0 to 3000 psi, or 0 to 5000 psi

 

● Atmospheric operating pressure: 11 to 29 psi

Summary of Contents for PSS 5000 Series

Page 1: ...ng of the gauge with other compatible devices Information that can be read from the TX Gauge includes device identity a record of events datalog the current values for configurable parameters and the firmware version Configurable parameters include intermediate pressure warning activation pressures gauge illumination duration etc The parameters described in this document are the default for the de...

Page 2: ...this may damage the equipment 1 Loosen the face mask straps As the seal between the face mask and the face is broken press the reset button Fig 3 Item 1 to switch off the positive pressure Fully remove the face mask and extend all of the straps of the head harness 2 Close the cylinder valve 3 Press the front button Fig 3 Item 2 to vent system and then press the reset button Fig 3 Item 1 to switch ...

Page 3: ...m the drying facility immediately when dry Drying time in a heated dryer must not exceed 30 minutes Do not immerse pneumatic or electronic components in cleaning solutions or water If water is trapped and then freezes inside the pneumatic system of the breathing apparatus such as the lung demand regulator its operation will be impaired Prevent any liquid from entering and thoroughly dry the breath...

Page 4: ...e and between multiple entries during the same use Decontamination and disposal procedures must be followed If contaminated with liquid chemical warfare agents dispose of the SCBA after decontamination U The respirator should not be used beyond 6 hours after initial exposure to chemical warfare agents to avoid possibility of agent permeation EBSS EBSS Activation or engagement of EBSS in either the...

Reviews: