background image

3362443 (A3-D-P) – Page 2 of 4

PSS 5000 Series

Self-contained breathing apparatus with Sentinel TX Gauge

Instructions for Use

i

3.1.2  Preparation for use

i

i

NOTICE

If the TX Gauge fails to switch on or a low battery alarm activates, 

renew the batteries.

1.  Carry out a visual inspection of the apparatus (see Section 3.5.1).

2.  Install the batteries if necessary (see Section 3.5.3).

3.  Pair the TX Gauge with one or more compatible PASS devices if 

required (see PASS device Instructions for Use).

4.  Fit the air cylinder (see Section 3.5.2).

5.  Press the male coupling of the lung demand regulator hose into the 

female coupling of the medium-pressure hose until an audible click is 

heard (do not connect the regulator to the face mask at this stage).

6.  Press the reset button (Fig 3, Item 1) to switch off the positive pressure. 

Press and rotate the bypass button (3) to align the red spots and then 

release the button to switch off the bypass.

7.  Carry out a full functional test of the apparatus (see Section 3.6).

8.  Align and push the lung demand regulator into face mask port until it 

latches in position, and check the attachment by gently attempting to 

pull the coupling apart.

3.2  Putting on the apparatus

1.  Fully loosen the shoulder straps and waist belt and put on the breathing 

apparatus.

2.  Check that the shoulder pads are not twisted and take the weight of 

the system on the shoulders by pulling the shoulder straps. Do not fully 

tighten at this stage.

3.  Close the waist belt buckle and pull the ends of the waist belt forward 

until  the  strap  padding  fits  securely  and  comfortably  over  the  hips 

(Fig 4). Tuck the belt ends behind the waist pad.

4.  Pull the shoulder straps until the breathing apparatus rests securely 

and comfortably on the hips. Do not over tighten. Pull the strap 

retainers down to secure the strap ends (Fig 5).

5.  Fully loosen the head straps of the face mask and place the neck strap 

over the back of the neck.

6.  Press the reset button (Fig 3, Item 1) to switch off the positive pressure.

7.  Open the cylinder valve (counterclockwise) slowly, but fully, to 

pressurize system. The TX Gauge switches on automatically when 

the cylinder valve is opened if the pressure in the air cylinder is 

approximately 200 psi or greater. The TX Gauge then performs a self 

check.

 

○ f the TX Gauge passes the self check the green LED will begin to 

flash once every second to indicate that it is switched on.

 

○ f the TX Gauge fails the self check during start up, or if a hardware 

failure occurs during use, the error alert activates (flashing of the 

amber LED at 2 second intervals for 10 seconds). Report the fault 

to trained service personnel. Do not use the breathing apparatus 

until the fault condition is rectified.

i

i

NOTICE

After storage at temperatures below 32 °F (0 °C) leakage may be 

observed when the cylinder  valve is initially opened due to ice 

formation.

 

● If leakage is observed from the 

lung demand regulator

Press the front button (Fig 3, Item  2) to allow a rush of air 

to pass through the lung demand regulator and then quickly 

press the  reset button (Fig 3, Item 1) to switch off the positive 

pressure. Resume normal operation.

 

● If leakage is observed from the 

quick  connect  cylinder 

coupling

: Close the cylinder valve  and vent the system. 

Disconnect then reconnect the cylinder to the breathing 

apparatus, then  reopen the cylinder valve slowly, but fully, to 

pressurize the system. Resume normal  operation.

 

● In the event that leakage still occurs, remove the breathing 

apparatus from service and  report the fault to trained service 

personnel or contact Dräger.

!

WARNING

For use in a CBRN environment, use only the face mask sizes that 

have been confirmed by a quantitative fit test (QNFT).

8.  Put on the face mask and check for tight fit (for non-CBRN use see 

the Dräger FPS 7000 Face Mask Instructions for Use; for CBRN 

use see the FPS NFPA Lung Demand Regulator and CBRN Special 

Instructions).

3.3  During use

!

WARNING

Fully open all cylinder valves and ensure that they remain open 

during use.

Users should be in a safe area before the end-of-service-time 

indicator (EOSTI) warning commences. Evacuate to a safe area 

immediately if the warning commences during an operation.

Using the bypass button (Fig 3, Item 3) will use air from the 

cylinder and rapidly reduce the working duration of the apparatus.

 

● Regularly check the remaining cylinder pressure on the gauge.

 

● Press the illumination button (Fig 2, Item 1) to illuminate the face of the 

TX Gauge with a white LED for approximately 3 seconds.

 

● If additional air is required, briefly press and release the bypass button 

(Fig 3, Item 3) to deliver a single jet of air into the face mask.

!

WARNING

The emergency air flow procedures below may greatly reduce the 

operating duration of the air supply. When activated the user must 

immediately evacuate to a safe area. The reason for using the 

procedure must be investigated and repaired before reusing the 

apparatus.

 

● Additional  air  flow  required  (emergency  procedure  only  used  in 

the  unlikely  condition  of  low  or  blocked  airflow)  –  Press  and  rotate 

the bypass button (Fig 3, Item 3) to deliver a sustained air supply 

(85 to 130 liters/minute) into the face mask.

 

● Excessive  or  loss  of  air  flow  (emergency  procedure  only  used  in 

the unlikely condition of high or loss of airflow) – Close the cylinder 

valve then immediately begin to slowly reopen the valve. Use the 

cylinder valve as a regulating valve to set the air flow to meet the user 

requirement. This procedure can be used with screw-type and ratchet-

type cylinder valves.

3.3.1  Low pressure warning

When the pressure in the air cylinder reaches a preset pressure range (see 

Section 8), the face of the TX Gauge will be illuminated by a flashing red 

LED and the low cylinder pressure warning whistle will sound.

i

i

NOTICE

The warning whistle and flashing red LED will activate within the 

preset pressure range. Due to the actuation tolerance, they may 

not activate at exactly the same time.

The red LED will switch off when the pressure in the air cylinder drops 

below 145 psi. The whistle will continue to sound until the cylinder pressure 

reaches zero.

3.3.2  Low battery warning

When the battery power reaches a low level, the amber LED on the 

TX Gauge will begin to flash once every 5 seconds. Replace the batteries 

at the earliest opportunity (see Section 3.5.3).

3.3.3  TX Gauge visual signals

The visual signals provided by the TX Gauge are shown in the table below.

Visual signal

Explanation

Flashing green LED

Gauge is switched on

Flashing red LED

Low cylinder pressure

Flashing amber LED once every 

5 seconds

Low battery

Flashing amber  LED at 2 second 

intervals for 10 seconds

Failed self check or hardware fault

Solid green LED

Attempting to establish 

communication with PC Link 

Module

Flashing green and amber LEDs

Communicating with PC Link 

Module

Solid green and white LEDs; 

flashing red LED

Attempting to pair with PASS 

device

Alternating flashing white and 

green/amber LEDs

Successfully paired with PASS 

device

Alternating flashing red and amber 

LEDs

Unable to pair with PASS device

3.4  After use

!

WARNING

Do not remove the breathing apparatus until in a safe breathing 

environment.

!

CAUTION

Do not remove the face mask by pulling on the lung demand 

regulator as this may damage the equipment.

1.  Loosen the face mask straps. As the seal between the face mask and 

the face is broken, press the reset button (Fig 3, Item 1) to switch off 

the positive pressure. Fully remove the face mask and extend all of the 

straps of the head harness.

2.  Close the cylinder valve.

3.  Press the front button (Fig 3, Item 2) to vent system and then press 

the reset button (Fig 3, Item 1) to switch off the positive pressure. The 

TX Gauge will switch off automatically when the pressure drops below 

145 psi.

4.  Release the waist belt buckle.

5.  Lift the shoulder strap ends to release the strap retainers (Fig  5) and 

then lift the shoulder strap buckles to loosen the straps.

6.  Remove the breathing apparatus and face mask.

7.  If the lung demand regulator has been set to bypass, press and rotate 

the bypass button (Fig 3, Item 3) to align the red spots and then 

release to switch off the bypass.

8.  Carry out the after use tasks in the maintenance table (Section  5.1).

9.  Remove the air cylinder if required (Section 3.5.3).

10. Pass the breathing apparatus to the service s department with details 

of any faults or damage that occurred during use.

3.5  Common user tasks

3.5.1  Visual inspection

A visual inspection must check the full breathing apparatus including all 

component parts and accessories. Check that the equipment is clean and 

undamaged, paying particular attention to pneumatic components, hoses 

and connectors. Typical signs of damage that may affect the operation of 

the breathing apparatus include impact, abrasion, cutting, corrosion and 

discoloration. Report damage to service personnel and do not use the 

apparatus until faults are rectified.

3.5.2  Air cylinder fitting and removing

!

WARNING

High-pressure air release may cause injury to the user or other 

personnel near the breathing apparatus. Close the cylinder valve 

and fully vent the system before attempting to disconnect an air 

cylinder.

Impact  damage  to  the  cylinder  valve  or  first-stage  regulator 

connector may prevent valve connection or cause an air leak. 

Handle the air cylinder and breathing apparatus with care.

i

i

NOTICE

The following instructions are for a threaded cylinder coupling. 

Fitting a quick-connect cylinder coupling is detailed in a separate 

document (see the Quick-Connect Cylinder Coupling User 

Instructions).

Fitting the cylinder

1.  Check  the  threads  of  the  cylinder  valve  port  and  the  first-stage 

regulator. Ensure that the O-ring seal (Fig 6, Item 1) in the first-stage 

regulator is clean and undamaged.

2.  Lay the backplate horizontally, with the first-stage regulator uppermost, 

and fully extend the cylinder strap.

3.  Insert the cylinder through the loop of the strap, and align the valve 

with the regulator.

4.  Lift the cylinder and backplate into the vertical position (supported on 

the end of the cylinder opposite the valve).

5.  Tighten the hand wheel of the regulator, using only the thumb and 

index finger, until a definite metal-to-metal contact is felt. Do not use 

tools or over tighten.

6.  Place the unit back into the horizontal position.

7.  Take up the slack in the cylinder strap (Fig 7).

8.  Pull the strap over the cylinder to operate the cam lock (Fig 8) and 

secure using the Velcro fastening.

Removing the cylinder

1.  Close the cylinder valve and press the front button (Fig 3, Item 2) to 

fully vent the system.

2.  Lay the backplate horizontally, with the cylinder uppermost.

3.  Remove the free end of the cylinder strap from the Velcro.

4.  Lift the strap against the cam lock to release the buckle tension and 

loosen the strap.

5.  Disconnect the cylinder valve from the first-stage regulator.

6.  Lift the cylinder away from the first-stage regulator and remove the 

cylinder.

3.5.3  Fitting or replacing the batteries

!

WARNING

Danger of explosion or fire. Do not remove or install the batteries 

in an explosive atmosphere.

Explosion, fire or chemical hazard. Do not expose the batteries to heat 

sources, do not attempt to recharge any non-rechargeable battery and 

do not short out the battery terminals.

Risk of explosion if a battery is replaced by an incorrect type. Use 

only the recommended battery type.

!

CAUTION

Environmental hazard. Dispose of used batteries in accordance 

with national or local regulations.

 

● Use only the recommended battery type: Panasonic CR123A 3 V 

lithium.

 

● Replace the batteries with a matching set. Do not mix new and used 

batteries.

 

● Remove discharged batteries from the product.

1.  Remove the battery cover using a 2.5 mm hexagon key, and remove 

the discharged batteries (see Fig 9).

2.  Note the polarity shown in the battery compartment and install the 

new batteries.

3.  Check the cover and sealing ring. Lightly coat the sealing ring with 

Dow Corning 111 silicone grease (as a guideline, the grease should be 

felt on the fingers but not seen).

4.  Align  and  refit  the  cover.  Gently  tighten  (nip  up)  the  screws.  Dräger 

recommend a torque of 0.4 lbf ft (0.5 Nm) – do not over tighten.

3.6  Functional testing

!

WARNING

Failure of the breathing apparatus to meet any of the standards 

or parameters described in the functional tests indicates a system 

fault. Report the fault to trained service personnel or contact 

Dräger. Do not use the breathing apparatus until the fault condition 

is rectified.

Assemble the breathing apparatus as described in the preparation for use 

(see Section 3.1) before commencing any functional testing.

Leak test and low pressure warning test

1.  Press the reset button (Fig 3, Item 1) to switch off the positive pressure. 

Press and rotate the bypass button (Fig 3, Item 3) to align the red spots 

and then release the button to switch off the bypass.

2.  Open the cylinder valve slowly, but fully, to pressurize the system. 

During pressurization a momentary sounding of the whistle will occur 

and the TX Gauge will switch on automatically.

3.  Fully close the cylinder valve.

4.  After 20 seconds, check the contents gauge and then reopen the 

cylinder valve. The gauge must not show an increase in pressure 

of more than 300 psi. If the pressure increase is more than 300 psi, 

investigate and repair the fault (see Section 4), and then repeat the 

leak test.

5.  Fully close the cylinder valve.

6.  Cover the outlet port of the lung demand regulator with the palm of 

the hand and press the front button (Fig 3, Item 2) to switch on the 

positive pressure.

7.  Carefully lift the palm of the hand to very slowly vent the system 

until the low pressure warning activates, and observe the pressure 

displayed on the gauge.

i

i

NOTICE

The warning whistle and flashing red LED will activate within the 

preset pressure range. Due to the actuation tolerance, they may 

not activate at exactly the same time.

8.  The low pressure warning must begin to sound in the range:

 

○ 216 psi cylinder: 819 psi to 732 psi

 

○  4500 psi cylinder: 1665 psi to 1485 psi

9.  Continue to vent the system until it is fully exhausted.

10. Press the reset button (Fig 3, Item 1) to switch off the positive pressure.

Troubleshooting

The troubleshooting guide shows fault diagnosis and repair information 

applicable to breathing apparatus users. Further troubleshooting and 

repair information is available in Instructions for Use supplied with 

associated equipment.

Where the troubleshooting guide shows more than one fault or remedy, 

carry out repair actions in the order that they appear in the table.

Contact service personnel or Dräger when the remedy information 

indicates a service task, or if the symptom remains after all remedy actions 

have been attempted.

Summary of Contents for PSS 5000 Series

Page 1: ...ng of the gauge with other compatible devices Information that can be read from the TX Gauge includes device identity a record of events datalog the current values for configurable parameters and the firmware version Configurable parameters include intermediate pressure warning activation pressures gauge illumination duration etc The parameters described in this document are the default for the de...

Page 2: ...this may damage the equipment 1 Loosen the face mask straps As the seal between the face mask and the face is broken press the reset button Fig 3 Item 1 to switch off the positive pressure Fully remove the face mask and extend all of the straps of the head harness 2 Close the cylinder valve 3 Press the front button Fig 3 Item 2 to vent system and then press the reset button Fig 3 Item 1 to switch ...

Page 3: ...m the drying facility immediately when dry Drying time in a heated dryer must not exceed 30 minutes Do not immerse pneumatic or electronic components in cleaning solutions or water If water is trapped and then freezes inside the pneumatic system of the breathing apparatus such as the lung demand regulator its operation will be impaired Prevent any liquid from entering and thoroughly dry the breath...

Page 4: ...e and between multiple entries during the same use Decontamination and disposal procedures must be followed If contaminated with liquid chemical warfare agents dispose of the SCBA after decontamination U The respirator should not be used beyond 6 hours after initial exposure to chemical warfare agents to avoid possibility of agent permeation EBSS EBSS Activation or engagement of EBSS in either the...

Reviews: