9023927 - 2nd edition - May 2007
Page 19 of 72 Pages
Start-up
Activating info mode
The info mode is used to display information about general unit settings and the
unit status.
●
Press and hold the »
« key (longer than 3 seconds) – information about the
units is displayed on several screens.
— Briefly press the »
« key to move to the next screen.
— Briefly press the »
« to move back to the previous screen.
— The info mode can be terminated at any time by pressing the »
« key.
— If no key is pressed for 30 seconds, the unit automatically returns to its
previous state.
Example of info mode:
Screen 1
Instrument information
Line 1 – Date and time
Line 2 – Software version
Line 3 – Unit Part No.
Line 4 – Unit Serial No.
Line 5 – Unit code
Screen2
Sensor information:
Line 1 – Sensor name
Line 2 – Sensor Part No.
Line 3 – Sensor Serial No.
Line 4 – EEPROM type
Line 5 – EEPROM version
Screen. 3
Sensor configuration:
Line 1 – Gas name
Line 2– Measuring range (cannot be change) and unit of measurement
Line 3 – Measuring range for the analogue interface.
displayed only if the analogue interface card is installed
Line 4 – A1 alarm threshold and unit of measurement
1)
Line 5 – A2 alarm threshold and unit of measurement
1)
1)
Displayed only if a relay module is fitted!
Screen 4
Pump Infos:
Line 1 – Pump flow
Line 2 – Threshold error
Line 3 – Threshold warning
Line 4 – Pump run time
0232392
7_1
.eps
M
OK
02
42392
7_1_en.eps
07.05.2005 12:34
SW Version: 1
Part No. : 8318827
Serial No. : ARUA0001
DeviceCode : 00006317
Instrument Info
1/4
0252392
7_1_en.eps
2/4
Sensorname: NF3
Part No.
: 6811125
Serial No. : XXXXXXX
EEPROM Typ
: 1
EEPROM Vers.: 1
Sensor Info
0262392
7_en.eps
3/4
Gasname : NF
3
Range : 30.0 ppm
4–20 SP
: 30.0 ppm
Alarm A1
: 5.0 ppm
Alarm A2
: 10.0 ppm
Sensor Config
02
7
2392
7_1_en.eps
4/4
Power
: XXX
%
Fault
: X.X
l/min
Warning : X.X
l/min
Op. time :
XXXX h
Pump Info