background image

8

Calibrating the gas detector

en

5

Calibrating the gas detector

WARNING

Danger to health!
Do not inhale the test gas. Observe the hazard warnings on 
the corresponding Safety Data Sheets and the instructions for 
use of the gas detector! Observe national regulations when 
defining calibration intervals.

For Pac 8500, two measuring channels are used for the 
calibration.

Calibration must be performed by qualified personnel after 
a failed bump test or after a defined calibration interval (see 
EU standard EN 50073).

Recommended calibration interval for the O

2

, H

2

S and CO 

sensors: 6 months. Calibration intervals for other gases: 
see the instructions for use of the respective 
DrägerSensors.

Calibration can be carried out as follows:

Manual calibration or calibration using the Dräger 
CC-Vision Basic

Automatic calibration with the Dräger X-dock (see the 
instructions for use for the Dräger X-dock)

Automatic calibration with the Dräger Bump Test Station 
(see the quick reference guide on the Dräger Bump Test 
Station)

5.1

Opening the maintenance menu

The gas detector automatically returns to measuring mode 
if no button is pressed in the menu for 1 minute (with the 
exception of the span calibration menus, where the menu 
remains active for 10 minutes without any activity).

The maintenance menu is password protected.
Factory setting: 001

1. Hold down the [

] button for 5 seconds.

2. Enter the password. Press the [

] button to change the 

value of the flashing position. Press the [OK] button to 
confirm the value. Repeat this procedure to define the next 
two values. After the last confirmation with the [OK] button, 
the entire password flashes.

3. Press the [OK] button to confirm the password or the [

button to cancel the password.

4. If the correct password is entered, a signal tone sounds 

and the fresh air calibration symbol flashes (configurable).

5. Press the [OK] button to open the fresh air calibration or the 

[

] button to switch to the span calibration.

6. Press the [OK] button to open the span calibration or the 

[

] button to switch back to measuring mode.

5.2

Carrying out a manual fresh air 
calibration

To improve accuracy, a fresh air calibration can be carried out 
if there is a zero deviation.

Observe the following notices for the calibration:

During fresh air calibrations, the zero point of all sensors 
(except the DrägerSensors XXS O

2

) is set to 0. For the 

XXS O

2

, the reading is set to 20.9 Vol%.

The DrägerSensor XXS O

3

 must be calibrated with a 

suitable zero gas which is free from carbon dioxide / ozone 
(e.g. N

2

).

To carry out the fresh air calibration:
1. Open the maintenance menu, see Section 5.1 on Page 8.
2. Press the [OK] button while the fresh air calibration symbol 

is flashing. The measured value flashes.

3. Press the [OK] button to confirm the fresh air calibration or 

the [

] button to cancel the fresh air calibration. 

The gas detector returns to measuring mode, or the span 
calibration symbol flashes (depending on the 
configuration).

Fresh air calibration passed: A short double tone sounds, 
and 

OK / gas name

 is shown in an alternating display. 

Press the [OK] button to return to measuring mode.

Fresh air calibration failed: A long single tone sounds.

– – –

 is displayed instead of the measured value. The error 

symbol and the fresh air calibration symbol are displayed. 
In this case, repeat the fresh air calibration or calibrate the 
gas detector.

Summary of Contents for Pac 6000

Page 1: ...Dr ger Pac 6x00 8x00 Technical manual i...

Page 2: ......

Page 3: ...ration 7 5 Calibrating the gas detector 8 5 1 Opening the maintenance menu 8 5 2 Carrying out a manual fresh air calibration 8 5 3 Carrying out a manual span calibration 9 5 4 Carrying out manual cali...

Page 4: ...and non earthed metal parts of the housing may store a combustible level of electrostatic charge Activities such as carrying the device in a bag or attached to a belt operating the button field or cle...

Page 5: ...ormity are provided in the enclosed supplementary documentation order no 90 33 741 The name plate on the gas detector must not be concealed 1 Alarm LEDs 6 OK button 2 Operation signal D Light 7 button...

Page 6: ...curacy After the gas detector has been switched on the current measured value is shown in the display Check whether the warning appears If it is displayed we recommend performing a bump test see Secti...

Page 7: ...vely The information display closes after the device ID When the gas detector is on The following information can be displayed depending on the configuration Error codes Peak concentration Time weight...

Page 8: ...the last confirmation with the OK button the entire password flashes 3 Press the OK button to confirm the password or the button to cancel the password 4 If the correct password is entered a signal to...

Page 9: ...r the button to cancel the span calibration Span calibration passed A short double tone sounds and OK gas name is shown in an alternating display Press the OK button to return to measuring mode Span c...

Page 10: ...Wrong software version Update the software 117 User parameters implausible Check and adjust the configuration of the user parameters 131 Gas detector faulty Contact the service of Dr ger 132 Paramete...

Page 11: ...rror on Dr gerSensor XXS measuring channel 2 Carry out a calibration see Section 5 on Page 8 327 Gas detector faulty Contact the service of Dr ger 328 Compensation channel calibration error on Dr gerS...

Page 12: ...n Page 8 351 Dr gerSensor XXS measuring channel 2 warming up warm up phase 1 Wait until warm up time is complete 352 Dr gerSensor XXS measuring channel 2 warming up warm up phase 2 Wait until warm up...

Page 13: ...ing the Dr ger CC Vision Basic PC software 1 Use the communication module order no 83 18 587 to connect the gas detector to a PC or the Dr ger X dock maintenance station 2 The configuration can be car...

Page 14: ...on Basic PC software as described Dr ger recommends signing on the sensor with the Dr ger CC Vision Basic PC software even if the sensor code is identical to the previous one 7 5 Replacing the battery...

Page 15: ...iaphragm must be replaced due to the special diaphragm 8 Device settings Only trained and qualified personnel may change the device settings 8 1 Factory setting The factory settings may differ for cus...

Page 16: ...Main alarm3 Instrument alarm 1 The intensity of the vibration depends on the temperature 2 After the initial battery pre alarm the battery s life span amounts to between 1 day and 2 weeks under normal...

Page 17: ...centers as specified by the applicable regulations Ambient conditions during operation Temperature temperature depending on the sensor Up to 30 C 55 C briefly up to 1 h 40 C 55 C Humidity 10 90 r h n...

Page 18: ...m Acknowledgeable Latching Alarm threshold A2 1 60 ppm 10 ppm 23 Vol 2 ppm Acknowledgeable Latching TWA threshold value 1 3 3 Not applicable for Pac 6000 30 ppm 10 ppm 1 ppm STEL threshold value 1 3 6...

Page 19: ...1 100 ppm 0 2 ppm 20 ppm 50 ppm 1 ppm 1 Vol Acknowledgeable Latching TWA threshold value 1 20 ppm 0 1 ppm 1 9 ppm 25 ppm 0 5 ppm 0 5 Vol STEL threshold value 1 40 ppm 0 1 ppm 3 8 ppm 50 ppm 1 ppm 2 V...

Page 20: ...A2 1 1 ppm 20 ppm 20 ppm 0 2 ppm 0 2 ppm Acknowledgeable Latching TWA threshold value 1 0 5 ppm 0 1 ppm 0 1 ppm STEL threshold value 1 0 5 ppm 0 1 ppm 0 1 ppm Mean value period 15 min 15 min 15 min Me...

Page 21: ...60 ppm CO 23 O 2 60 CO 60 ppm Acknowledgeable O 2 CO Latching O 2 CO TWA threshold value 1 5 ppm H 2 S 30 ppm CO 30 ppm CO 30 ppm CO STEL threshold value 1 10 ppm H 2 S 60 ppm CO 60 ppm CO 60 ppm CO M...

Page 22: ...2 No effect Ethane No effect No effect 0 2 Ethanol No effect No effect No effect Ethylene No effect No effect 1 Hydrogen 0 35 No effect 1 5 Hydrogen chloride No effect No effect No effect Hydrogen cy...

Page 23: ......

Page 24: ...ger Safety AG Co KGaA Revalstra e 1 23560 L beck Germany Tel 49 451 882 0 Fax 49 451 882 20 80 www draeger com 90 33 742 TH 4623 700 Dr ger Safety AG Co KGaA Edition 01 July 2017 Subject to alteratio...

Reviews: