Device settings
15
en
7.6
Replacing the dust and water filter
1. Place the special tool on the dust and water separator.
2. Press the lever on the special tool downward and then
remove the dust and water separator.
3. Insert a new dust and water separator. The dust and water
separator must click into place.
For special gases (ozone, phosgene), the entire front casing,
including the special diaphragm, must be replaced due to the
special diaphragm.
8
Device settings
Only trained and qualified personnel may change the device
settings.
8.1
Factory setting
The factory settings may differ for customised orders.
Sensor grille:
4 units
Order no.:
40 units
Order no.:
Pac 6xx0
83 26 853
83 26 857
Pac 8xx0
83 26 852
83 26 859
Front casing:
Order no.:
Pac 8000 ozone
83 26 851
Pac 8000 phosgene
83 26 854
00633742.eps
Bump test mode
Quick bump test
Vibration alarm
On
Bump test interval
Off
Operation signal
On
D-Light
On
Switch off
Always
Data logger interval
1 min
Usage duration (user)
Off