Dräger Pac 6000 Technical Manual Download Page 14

14

Maintenance

en

If the gas detector can no longer be calibrated, the sensor 
must be replaced!

Only use the DrägerSensor XXS with the same part 
number!

1. Connect the gas detector to a PC using the communication 

module.

2. In the CC-Vision Basic PC software

1)

, open the “Sensor 

replacement assistant” feature and follow the instructions.
a. Switch off the gas detector.
b. Loosen the four screws on the rear part of the housing.
c. Open the front part of the housing.
d. Remove the battery (optional).
e. Remove the sensor.
f. Insert the new sensor and make a note of the sensor 

code printed on the sensor.

g. If the battery was removed: Wait for 5 seconds, insert 

new battery. The battery runtime is reset when the 
battery is inserted.

h. Place the front part of the housing on the gas detector 

and re-tighten the four screws on the rear part of the 
housing.

3. Sign in the sensor with the previously recorded sensor 

code.

4. After the sensor replacement, the sensor requires a warm-

up phase (see the sensor data sheet). The displayed 
concentration flashes until the warm-up phase is complete.

5. After the sensor replacement and the warm-up phase, the 

gas detector must be calibrated (see Section 5 on Page 8).

If the sensor code of the new sensor does not match that 
of the previous sensor, the new sensor must be signed on 
with the CC-Vision Basic PC software as described. Dräger 
recommends signing on the sensor with the Dräger 
CC-Vision Basic PC software, even if the sensor code is 
identical to the previous one.

7.5

Replacing the battery

WARNING

Risk of explosion! 

Only the lithium battery type LBT 01** (order no. 
83 26 856) may be used.

Do not remove or replace batteries in potentially explosive 
areas. Do no throw used batteries into fire or try to open 
them by force. Dispose of batteries in accordance with the 
national provisions.

The battery is part of the Ex approval.

1. Switch off the gas detector.
2. Loosen the four screws on the rear part of the housing.
3. Open the front part of the housing, remove battery and wait 

for 5 s.

4. Insert the new battery. Make sure the battery plug is 

inserted correctly (see figure). The battery runtime is reset 
when the battery is inserted.

5. Place the front part of the housing on the gas detector and 

re-tighten the four screws on the rear part of the housing 
(torque: 35 Ncm).

6. Turn on the gas detector.

A successful battery replacement is confirmed with a 
vibration (5 s) when the gas detector is turned on.
Notice 160 is displayed after the start sequence (see 
Section 6 on Page 10).

7. Confirm the notice with the [OK] button.
8. After the battery replacement, the sensor requires a warm-

up phase (see the sensor data sheet). The displayed 
concentration flashes until the warm-up phase is complete.

1) A free version of the Dräger CC-Vision Basic PC software can be 

downloaded at www.draeger.com/software

35 Ncm

00533742.eps

Summary of Contents for Pac 6000

Page 1: ...Dr ger Pac 6x00 8x00 Technical manual i...

Page 2: ......

Page 3: ...ration 7 5 Calibrating the gas detector 8 5 1 Opening the maintenance menu 8 5 2 Carrying out a manual fresh air calibration 8 5 3 Carrying out a manual span calibration 9 5 4 Carrying out manual cali...

Page 4: ...and non earthed metal parts of the housing may store a combustible level of electrostatic charge Activities such as carrying the device in a bag or attached to a belt operating the button field or cle...

Page 5: ...ormity are provided in the enclosed supplementary documentation order no 90 33 741 The name plate on the gas detector must not be concealed 1 Alarm LEDs 6 OK button 2 Operation signal D Light 7 button...

Page 6: ...curacy After the gas detector has been switched on the current measured value is shown in the display Check whether the warning appears If it is displayed we recommend performing a bump test see Secti...

Page 7: ...vely The information display closes after the device ID When the gas detector is on The following information can be displayed depending on the configuration Error codes Peak concentration Time weight...

Page 8: ...the last confirmation with the OK button the entire password flashes 3 Press the OK button to confirm the password or the button to cancel the password 4 If the correct password is entered a signal to...

Page 9: ...r the button to cancel the span calibration Span calibration passed A short double tone sounds and OK gas name is shown in an alternating display Press the OK button to return to measuring mode Span c...

Page 10: ...Wrong software version Update the software 117 User parameters implausible Check and adjust the configuration of the user parameters 131 Gas detector faulty Contact the service of Dr ger 132 Paramete...

Page 11: ...rror on Dr gerSensor XXS measuring channel 2 Carry out a calibration see Section 5 on Page 8 327 Gas detector faulty Contact the service of Dr ger 328 Compensation channel calibration error on Dr gerS...

Page 12: ...n Page 8 351 Dr gerSensor XXS measuring channel 2 warming up warm up phase 1 Wait until warm up time is complete 352 Dr gerSensor XXS measuring channel 2 warming up warm up phase 2 Wait until warm up...

Page 13: ...ing the Dr ger CC Vision Basic PC software 1 Use the communication module order no 83 18 587 to connect the gas detector to a PC or the Dr ger X dock maintenance station 2 The configuration can be car...

Page 14: ...on Basic PC software as described Dr ger recommends signing on the sensor with the Dr ger CC Vision Basic PC software even if the sensor code is identical to the previous one 7 5 Replacing the battery...

Page 15: ...iaphragm must be replaced due to the special diaphragm 8 Device settings Only trained and qualified personnel may change the device settings 8 1 Factory setting The factory settings may differ for cus...

Page 16: ...Main alarm3 Instrument alarm 1 The intensity of the vibration depends on the temperature 2 After the initial battery pre alarm the battery s life span amounts to between 1 day and 2 weeks under normal...

Page 17: ...centers as specified by the applicable regulations Ambient conditions during operation Temperature temperature depending on the sensor Up to 30 C 55 C briefly up to 1 h 40 C 55 C Humidity 10 90 r h n...

Page 18: ...m Acknowledgeable Latching Alarm threshold A2 1 60 ppm 10 ppm 23 Vol 2 ppm Acknowledgeable Latching TWA threshold value 1 3 3 Not applicable for Pac 6000 30 ppm 10 ppm 1 ppm STEL threshold value 1 3 6...

Page 19: ...1 100 ppm 0 2 ppm 20 ppm 50 ppm 1 ppm 1 Vol Acknowledgeable Latching TWA threshold value 1 20 ppm 0 1 ppm 1 9 ppm 25 ppm 0 5 ppm 0 5 Vol STEL threshold value 1 40 ppm 0 1 ppm 3 8 ppm 50 ppm 1 ppm 2 V...

Page 20: ...A2 1 1 ppm 20 ppm 20 ppm 0 2 ppm 0 2 ppm Acknowledgeable Latching TWA threshold value 1 0 5 ppm 0 1 ppm 0 1 ppm STEL threshold value 1 0 5 ppm 0 1 ppm 0 1 ppm Mean value period 15 min 15 min 15 min Me...

Page 21: ...60 ppm CO 23 O 2 60 CO 60 ppm Acknowledgeable O 2 CO Latching O 2 CO TWA threshold value 1 5 ppm H 2 S 30 ppm CO 30 ppm CO 30 ppm CO STEL threshold value 1 10 ppm H 2 S 60 ppm CO 60 ppm CO 60 ppm CO M...

Page 22: ...2 No effect Ethane No effect No effect 0 2 Ethanol No effect No effect No effect Ethylene No effect No effect 1 Hydrogen 0 35 No effect 1 5 Hydrogen chloride No effect No effect No effect Hydrogen cy...

Page 23: ......

Page 24: ...ger Safety AG Co KGaA Revalstra e 1 23560 L beck Germany Tel 49 451 882 0 Fax 49 451 882 20 80 www draeger com 90 33 742 TH 4623 700 Dr ger Safety AG Co KGaA Edition 01 July 2017 Subject to alteratio...

Reviews: