
For your safety and that of your patients
6
Instructions for use Movita
Accessories
Connected devices
Safe connection with other electrical
equipment
Connection to other devices
Patient safety
The design of the medical device, the
accompanying documentation, and the labeling on
the medical device are based on the assumption
that the purchase and the use of the medical device
are restricted to persons familiar with the most
important inherent characteristics of the medical
device.
Instructions and WARNING and CAUTION
statements are therefore largely limited to the
specifics of the Dräger medical device.
The instructions for use do not contain any
information on the following points:
– Risks that are obvious to users
– Consequences of obvious improper use of the
medical device
– Potentially negative effects on patients with
different underlying diseases
Medical device modification or misuse can be
dangerous.
WARNING
Risk due to incompatible accessories
Dräger has tested only the compatibility of
accessories listed in the current list of
accessories G15600. If other, incompatible
accessories are used, there is a risk of patient
injury due to medical device failure.
Dräger recommends that the medical device is
only used together with accessories listed in
the current list of accessories.
WARNING
Risk of electric shock and of device
malfunction
Any connected devices or device
combinations not complying with the
requirements mentioned in these instructions
for use can compromise the correct
functioning of the medical device and lead to
an electric shock. Before operating the
medical device, strictly comply with the
instructions for use of all connected devices
or device combinations.
WARNING
Risk of patient injury
Electrical connections to equipment not listed
in these instructions for use or these
assembly instructions must only be made
when approved by each respective
manufacturer.
WARNING
Device combinations approved by Dräger
meet the requirements of the following
standards (if applicable):
– IEC 60601-1
Medical electrical equipment
Part 1: General requirements for safety
– IEC 60601-1-2
Medical electrical equipment
Part 1-2: General requirements for safety
Collateral standard: Electromagnetic
compatibility; Requirements and tests