34
Pocket Guide
Evita XL
Cleaning and disinfection
Cleaning and disinfection
NOTE
The
Evita XL
Pocket Guide is not a replacement or substitute for the Instructions for Use.
Any use of the device requires full understanding and strict observation of the Instructions for Use.
Components which can be
reprocessed
Recommended
reprocessing
intervals
Machine
cleaning and
disinfection
Manual
Sterilization
Cleaning Disinfection
Evita XL
basic device
after each patient
no
outside
outside
no
Trolley
after each patient
no
outside
outside
no
Hinged arm
Gas supply hose
Breathing hoses
after each patient/
weekly
yes
possible
possible
yes
Y-piece
Water traps
Collection containers
Adapter parts for the medication
nebulizer
Reusable expiratory valve and, if
necessary, individual parts
after each patient/
weekly
1)
yes
possible
possible
2)
yes
Disposable expiratory valve
Dispose of after each patient/weekly
Spirolog flow sensor
daily
no
outside
yes
no
SpiroLife flow sensor
daily
no
outside
yes
yes
Neonatal flow sensor insert
daily
no
outside
yes
yes
Neonatal flow sensor housing
daily
yes
possible
possible
yes
Temperature sensor
daily
no
outside
outside
3)
yes
CO
2
sensor
daily
no
outside
outside
4)
no
Reusable cuvette of the CO
2
sen-
sor
daily
yes
possible
possible
yes
Disposable cuvette of the CO
2
sensor
Dispose of after each patient/weekly
Test filter for CO
2
sensor
daily
no
outside
4)
outside
no
Breathing gas humidifier
after each patient/
weekly
In accordance with separate Instructions for Use
Medication nebulizer
In accordance with separate Instructions for Use
Bacterial filter
In accordance with separate Instructions for Use
1)
Nebulizing may lead to formation of more extensive deposits requiring more frequent replacement.
2)
Additionally treat the expiratory valve after disinfection.
3)
Do not bath-disinfect.
4)
Do not bath-disinfect. Wipe-disinfect, e. g., with 70 % ethanol.