Preparation
Installing expiratory valve, flow
sensor
Installing the expiratory valve
1
Press down the segments on the right and left and
tilt the control panel (A) upwards.
2
Push expiratory valve (B) into the mount until it
clicks into position. Check that it is properly engaged
by gently pulling on the port.
Mounting the flow sensor
1
Push connector socket (A) all the way to the left.
2
Place flow sensor (B) into its mount – with the con-
nector facing towards the device – and push into the
socket, as far as it will go.
3
Push flow sensor to the right and into the rubber lip
seal of the expiratory valve, as far as it will go.
Connecting the breathing circuit
1
Attach hinged arm (A) to the rail on the left-hand
side of the ventilator and tighten screws.
2
Connect breathing hoses of appropriate lengths to
the ports (B). Observe the required hose lengths (in-
dicated in meters).
3
Install water traps (C) in vertical position.
4
Connect Y-piece (D), with the rubber sleeve of the
Y-piece on the inspiratory side.
5
Insert the Y-piece in the opening of the hinged arm
(E).
6
Connect test lung.
007
008
009
A
B
A
B
501
0.6m
0.6m
0.4m
1.1m
D
Evita XL
A
B
B
C
D
E