![Dräger EVITA XL Pocket Manual Download Page 20](http://html.mh-extra.com/html/drager/evita-xl/evita-xl_pocket-manual_2526756020.webp)
20
Pocket Guide
Evita XL
Operation
Overview of the availability of additional
settings
Bronchial suctioning
Before suctioning
1
Touch
Special Procedure ...
button in the main
menu bar.
The
Additional Function
page (A) appears by default.
2
Touch
O
2
↑
suction
button (B).
3
Press rotary knob to confirm.
The oxygenation procedure is started.
Evita XL
ventilates the patient in the set ventilation
mode with the appropriately increased O
2
concentrati-
on.
The preoxygenation phase with the remaining time in
seconds is displayed continuously in the header bar.
Preoxygenation lasts for a maximum of 180 seconds.
During this time,
Evita XL
waits for the disconnection
necessary for suctioning. If no disconnection is detected
within 180 seconds,
Evita XL
terminates the oxygenati-
on procedure.
During suctioning
The time available for suctioning is displayed in se-
conds in the header bar. If suctioning is completed and
the patient reconnected within the time available,
Evita XL
will end the disconnection phase.
After suctioning
After reconnection,
Evita XL
resumes ventilation in the
set ventilation mode. For postoxygenation, the O
2
con-
centration is increased for the first 120 seconds.
The postoxygenation phase with the remaining time in
seconds is displayed in the header bar.
Canceling oxygenation
Touch
O
2
↑
suction
button.
Ventilation
mode
Additional settings
ATC
(optional)
Apnea venti-
lation
Flowtrigger
Inspiratory
termination
AutoFlow
Sigh
PLV
CMV
X
X
X
X
X
SIMV
X
X
X
X
X
X
MMV
X
X
X
X
X
ILV
Master
X
X
X
X
ILV
Slave
X
X
X
PCV+
X
X
X
X
PCV+ Assist
X
X
APRV
X
X
CPAP/ PSupp
X
X
X
X
PPS (optional)
X
X
X
026
A
B