Dräger Alcotest 6820 med
Use
3
Use
3.1
Prerequisites
The ambient air must be free from alcohol and solvent vapours
(disinfectants).
Maintain an adequate distance from mobile telephone antennas
and transmitter sites so as not to impair the function of the device.
Observe a waiting period of at least 15 minutes after the last oral
alcohol intake!
Residual alcohol in the mouth can distort the measurement. Even
aromatic beverages (e. g. fruit juice), alcoholic mouth sprays,
medical juices and drops and, belching and vomiting may result in
false data. Also observe a waiting period of at least 15 minutes in
these cases. Rinsing the mouth with water or non-alcoholic drinks
does not replace the waiting time!
The test person should breathe regularly and calmly before the
sample is taken. Avoid breathing in and out quickly through the
mouth.
The device records breath samples with a volume greater than 0.3
L. The test subject must be able to deliver this volume as a
minimum. If the volume of the breath sample is over 0.3 L and less
than 1.2 L, the message "Low volume" is shown on the display as
well as the measurement result.
The displayed alcohol content of the breath sample does not
reflect the blood alcohol concentration until the volume exceeds
1.2 L breath volume. All alcohol tests of breath samples with the
message "Low volume" may display a lower result than the actual
blood alcohol concentration.
3.2
Preparations for use
3.2.1
Inserting or replacing the batteries / rechargeable
batteries
1.
Undo the screw plug on the battery lid (e. g. using a coin) and
remove the battery lid.
2.
Insert the batteries or rechargeable batteries and make sure that
the polarity is correct.
3.
Close the battery lid and tighten the screw plug.
Required batteries/rechargeable batteries:
2 x 1.5 V alkaline type (AAA, LR6, AA) or,
2 rechargeable batteries - 1.2 V NiMH (AAA, LR6, AA)
CAUTION
Explosion hazard!
Do not throw used batteries/rechargeable batteries into a fire
or open them by force. Do not recharge alkaline batteries.
Batteries/rechargeable batteries must be disposed of in
accordance with local and national regulations.
NOTICE
To ensure that the device can be used again immediately
after the alkaline batteries or the rechargeable NiMH batteries
are discharged, it is recommended to have replacement
batteries available.
Replace the batteries within 5 minutes, otherwise the save
time will be reset and must be set again.
!