Reprocessing Fabius family
13
English
Removing the flow sensor
Observe the instructions for use of the flow sensor.
1
Loosen the expiratory port (A) and pull it out.
2
Remove the flow-sensor guard (B) (option).
3
Remove the flow sensor (C).
4
Refit the expiratory port (A).
Removing the compact breathing system
1
Completely pull out the locking pin (A) and hold
it in that position.
2
Remove the compact breathing system from
the holder (B) by lifting it upwards.
3
Remove the compact breathing system cover
(option, not shown).
140
A
B
C
CAUTION
Risk of injury due to breathing system heating
When the breathing system heating is switched
on, the bottom side of the compact breathing sys-
tem and the heating plate beneath it can become
hot.
Allow the compact breathing system to cool off
before removing it.
141
NOTE
To prevent accidental penetration of soda lime
into the breathing system, do not transport the
compact breathing system with a filled reusable
absorber.
A
B