![Dräger 9054695 Reprocessing Manual Download Page 11](http://html1.mh-extra.com/html/drager/9054695/9054695_reprocessing-manual_2526447011.webp)
Reprocessing Fabius family
11
English
Removing the CO
2
absorber
1
Remove the CO
2
absorber:
– CLIC absorber (disposable absorber)
Or
– Reusable absorber
CLIC absorber (disposable absorber)
Observe the instructions for use of the CLIC
absorber.
1
Press the release button (A). The CLIC adapter
flips open.
2
Pull the CLIC absorber (B) upwards out of the
holder (C).
3
Flip the holder back until it audibly clicks into
place.
Reusable absorber
1
Unscrew the CO
2
absorber (A) from the
compact breathing system.
2
Remove and dispose of the disposable dust
filter (B) (option).
3
Empty the used soda lime. Dispose of the soda
lime according to its instructions for use.
4
Remove the absorber insert (C) from the
absorber container. The sealing ring (D)
remains on the absorber insert.
076
A
C
B
128
013
A
B
D
C