![Dräger Medical Savina Manual Download Page 90](http://html.mh-extra.com/html/drager-medical-0/savina/savina_manual_2527053090.webp)
Page 4 of 4
Recommended separation distances
Recommended separation distances between portable and mobile RF telecommunication
devices and the Savina
max.
P
EIRP
(W)
3 V/m
distance*
(m)
1 V/m
distance*
(m)
Note
0.001 0.06 0.17
0.003 0.10 0.30
0.010 0.18 0.55
0.030
0.32
0.95
e.g. WLAN 5250 / 5775 (Europe)
0.100
0.58
1.73
e.g. WLAN 2440 (Europe), Bluetooth
0.200
0.82
2.46
e.g. WLAN 5250 (not in Europe)
0.250
0.91
2.75
e.g. DECT devices
1.000 1.83 5.48
e.g. GSM 1800- / GSM 1900- / UMTS- mobiles,
WLAN 5600 (not in Europe)
2.000
2.60
7.78
e.g. GSM 900 mobiles
3.000 3.16 9.49
Information re separation distances (IEC 60601-1-2: 2001, tables 205 and 206)
* 3 V/m distance to transmitters with frequencies from 150 kHz to 2.5 GHz, otherwise 1 V/m distance
.