5664.900 Savina
Version 4.0
Test List
A
ll right
s reserved. Co
pyright
reserved.
K
5
6
649
00
TL1.
fm
16
.12
.05
4
5
Accompanying documents check
Instructions for Use manual is available
Medical product logbook is available (applies to Germany only)
6
Safety checks
•
Check the power cable for possible damage.
•
The fuse links are as specified on the rear panel.
6.1
Test of electrical safety to VDE 0751
6.1.1
Protective earth conductor resistance
•
Measure the protective earth conductor resistance according to VDE 0751.
The protective conductor resistance should be less than/equal to 0.3 ohms.
6.1.2
Equivalent unit leakage current
•
Measure the equivalent device leakage current according to VDE 0751.
The equivalent leakage current is less than/equal to 1000
P
A
6.1.3
Equivalent patient leakage current
•
Measure the equivalent patient leakage current according to VDE 0751.
The equivalent patient leakage current is less than/equal to 5000
P
A
Summary of Contents for Savina
Page 2: ......
Page 6: ...All rights reserved Copyright reserved ENK5664900IECIVZ fm 04 01 06 IV Contents ...
Page 7: ...1 General ...
Page 8: ...2 ...
Page 11: ...5 Function Description ...
Page 12: ...6 ...
Page 29: ...23 Maintenance Procedures ...
Page 30: ...24 ...
Page 45: ...39 Schematics and Diagrams ...
Page 46: ...40 ...
Page 63: ...Annex Parts catalog Test List Technical Information ...
Page 64: ......
Page 66: ...2 ...
Page 91: ......