Maintenance procedures
1
General notes
For “replacement of wear and tear parts” intervals, refer to the Instructions for
Use manual. The following replacement procedures are described in this doc-
ument:
–
Replacing the microfilter
–
Replacing the dust filter set
–
Replacing the O2 sensors
–
Replacing the filter in the O2 inlet
–
Replacing the lead-gel battery (internal battery)
Read and observe the following warnings before opening the device:
2
Replacing the micro-
filter
1. Press and hold the two tabs (
Figure 1
-1).
2. Lift and remove the filter cover (
Figure 1
-2).
Figure 1
Removing the filter cover
3. Pull the contaminated microfilter out of its mount (
Figure 2
-1).
Disinfect and clean the unit or parts of the unit prior to each mainte-
nance procedure.
Hazardous voltage. Touching live components can lead to serious
injury or death. Disconnect power cord from AC outlet before opening
device.
Electrostatic discharge may damage electrostatic sensitive devices.
When handling electrostatic sensitive devices, use a static-dissipa-
tive mat and a static dissipative wrist strap.
Summary of Contents for Savina
Page 2: ......
Page 6: ...All rights reserved Copyright reserved ENK5664900IECIVZ fm 04 01 06 IV Contents ...
Page 7: ...1 General ...
Page 8: ...2 ...
Page 11: ...5 Function Description ...
Page 12: ...6 ...
Page 29: ...23 Maintenance Procedures ...
Page 30: ...24 ...
Page 45: ...39 Schematics and Diagrams ...
Page 46: ...40 ...
Page 63: ...Annex Parts catalog Test List Technical Information ...
Page 64: ......
Page 66: ...2 ...
Page 91: ......