32
6132.300
Function Description
Babytherm 8004/8010
Figure 17
Control unit (rear view), cover plate removed
Legend
The control unit contains the following assemblies:
–
ON/OFF switch
–
Toroidal-core power transformer
–
WT Power PCB
–
WT Relay PCB
–
WT Controller PCB
–
Front (WT LED 8010 PCB, membrane keypad 8010, text display,
WT Front PCB, loudspeaker)
2.2.1
ON/OFF switch
The ON/OFF switch is used to switch the mains voltage (single-pole) ON or
OFF.
1
WT Sensor PCB socket
2
Maximum of 3 connections for auxiliary equipment (workplace),
nurse call (connection for central alarm)
3
Connection for internal paging system, nurse call (connection for
central alarm)
4
RS 232 interface (modem), option
5
RS 232 interface (BabyLink), option
6
Control unit ON/OFF switch
7
Mains voltage selector
8
Fuses
9
Radiant warmer connection
10
Fuses
11
Mattress heater connection
12
Mains voltage supply connection