Function Description
Babytherm 8004/8010
Control and feedback of the infrared elements
An optocoupler controls the triac which supplies the infrared elements with
mains voltage. This current flow generates a feedback signal via the optocou-
pler.
Safety relay of the infrared elements
In the event of a malfunction (e.g. “Inop”), the safety relay switches off the
mains voltage to the infrared elements.
Relay for the energy-saving lamps and phototheray assembly (option)
The relays supply the energy-saving lamps and the the phototheray assem-
bly (option) with mains voltage.
Control of the Central Alarm PCB
An opto-triac on the WT Relay PCB controls the central alarm light. The
forced-tripped opto-triac supplies the Central Alarm PCB with mains voltage.
1.2.5
WT Controller PCB
The WT Controller PCB carries the following subassemblies:
–
Microcontroller
–
Quartz
–
EPROM
–
EEPROM
–
RAM
–
Watchdogs
–
Driver modules
–
Multiplexer and A/D converter
Summary of Contents for Babytherm 8004