This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections
and maintenance work carried out by the manufacturer.
6132.300 Babytherm 8004/8010
Test List
Page 8
K6132300TL1.fm
24.03.06
Copyright
reserved. Rep
roduct
ion
allow
e
d f
o
r non-commer
c
ia
l purposes only.
5
Functional Tests
5.1
Checking the LEDs, the LED Strip and the Buz-
zer in the Control Unit as well as Central Alarm
Lamp in the Radiant Heater
•
Plug the power cord of the Babytherm 8004 into the mains power
outlet.
•
Switch on the control unit.
•
Press the "Check" key.
The LEDs and the LED strip on the control unit come on. The central
alarm lamp in the radiant heater comes on. The buzzer in the control
unit sounds.
5.1.1
Checking the Infrared Rods in the Manual Mode
•
Switch on the control unit.
•
Set "manual mode" on control unit.
The infrared rods heat up.
5.1.2
Checking the Infrared Rods in the Skin Mode
•
Connect skin-temperature sensors to the Babytherm.
•
Switch on the control unit.
•
Set "skin mode" on control unit.
•
Place fingers firmly around ends of skin-temperature sensor
cables.
The infrared rods heat up. The temperature displayed on the control
unit increase.
Summary of Contents for Babytherm 8004
Page 2: ......
Page 7: ...1 General ...
Page 8: ...2 ...
Page 11: ...5 Function Description ...
Page 12: ...6 ...
Page 63: ...57 Maintenance Procedures ...
Page 64: ...58 ...
Page 73: ...Annex Parts catalog Test List Technical Information ...
Page 74: ......
Page 76: ...2 ...
Page 108: ......
Page 109: ......