background image

6.3

 

Fitting the cuff 

Insert air connector into air outlet shown in left photo and please make 
sure the fitting of the air connector completely and properly to avoid air 
leakage.   

a)

 

The distance between the edge of cuff and the elbow should be 
approx. 2~3cm.   

b)  Secure  the  cuff  with  the  Velcro  fastener,  so  that  it  lies 
comfortably and not too tight, whereby no space should remain 
between the cuff and the arm.   

c) Lay the arm on a table, with the palm upwards. Support the arm 
a little with a rest (cushion), so that the cuff rests at about the same 
height as the heart. Take care, that the cuff lies free. Remain so for 
2 minutes sitting quietly, before beginning with the measurement.   

d) Let legs uncrossed, feet flat on the floor, back and arm supported. 

Summary of Contents for Afib

Page 1: ...Afib Talk Blood Pressure Monitor Instruction Manual ...

Page 2: ...HE SET DATE 5 3 LANGUAGE SETTING USER SELECTION AND SETTING THE TIME DATE 6 Carrying Out A Measurement 6 1 BEFORE THE MEASUREMENT 6 2 COMMON SOURCES OF ERROR 6 3 FITTING THE CUFF 6 4 MEASURING PROCEDURE 6 5 DISCONTINUING A MEASUREMENT 6 6 MEMORY STORAGE AND RECALL OF THE MEASUREMENTS 6 7 MEMORY CANCELLATION OF ALL MEASUREMENTS ATTENTION 7 Appearance Of IHB Irregular Heart Beat Indicator For Early ...

Page 3: ...ion of a different component might result in measurement error Cuff is replaceable only by an original Do not use with neonatal patients Do not intend to use with pregnant or pre eclamptic patients It will cause harmful injury to the patient or affect the blood pressure due to connection tubing kinking Too frequent measurements can cause injury to the patient due to blood flow interference The app...

Page 4: ...surement 2 1 How does high low blood pressure arise As your heart beats it pumps your blood round your body so that your muscles can get all the energy and oxygen they need To do this your heart pushes your blood through a network of blood vessels called arteries As the blood travels through the arteries it pushes against the sides of these blood vessels and the strength of this pushing is called ...

Page 5: ...cific times of the day Show these values to your doctor Never use the results of your measurements to alter independently the drug doses prescribed by your doctor Table for classifying blood pressure values unit mmHg according to World Health Organization Range Systolic Blood pressure Diastolic Blood pressure Measures Hypotension lower than 100 lower than 60 Consult your doctor Blood pressure opti...

Page 6: ... the chance of having a stroke increases with age AFIB screening is recommended for people over 65 years and older However for people from the age of 50 years with high blood pressure hypertension diabetes coronary heart failure or have had a previous stroke AFIB screening is also recommended Early diagnosis of AFIB followed by adequate treatment can significantly reduce the risk of getting stroke...

Page 7: ...4 Various components of the blood pressure monitor ...

Page 8: ...e use AA Long Life or Alkaline1 5V Batteries The use of 1 2V Accumulators is not recommended If the blood pressure monitor is left unused for long periods please remove the batteries from the device 5 2 Reading the set date Please press the TIME button and the date will be shown in the display 5 3 Language selection User selection and setting the time date Talking Language selection Bilingual of E...

Page 9: ...s firstly user icon will blink Then press TIME button again the display now indicates the set year during which the four characters blink 2 The correct year can be entered by pressing the MEMORY button 3 Press the TIME button again The display now switches to the current date during which the first character month blinks 4 The corresponding month can now be entered by pressing the MEMORY button 5 ...

Page 10: ...the blood pressure changes during the course of the day 6 2 Common sources of error Note Comparable blood pressure measurements always require the same conditions These are normally always quiet conditions All efforts by the patient to support the arm can increase the blood pressure Make sure you are in a comfortable relaxed position and do not activate any of the muscles in the measurement arm du...

Page 11: ... cuff with the Velcro fastener so that it lies comfortably and not too tight whereby no space should remain between the cuff and the arm c Lay the arm on a table with the palm upwards Support the arm a little with a rest cushion so that the cuff rests at about the same height as the heart Take care that the cuff lies free Remain so for 2 minutes sitting quietly before beginning with the measuremen...

Page 12: ...ection during measuring the icon will appear if a movement was detected which may influence accuracy due to the movement not too serious the measuring can be continuous if the movement is too serious Err2 displayed d After reaching the inflation pressure the pump stops and the pressure slowly falls away The cuff pressure large characters is displayed during the measurement When the device has dete...

Page 13: ...es 6 4 2Measuring in Afib Mode 2 measurement mode In Afib mode 2 measurements are automatically taken in succession and the result is then automatically analyzed and displayed Because blood pressure constantly fluctuates a result determined in this way is more reliable than one produced by a single measurement After pressing the button the symbol appears in the display The middle left hand section...

Page 14: ...le a third one is automatically taken In the measuring After reaching the inflation pressure the pump stops and the pressure slowly falls away The cuff pressure is displayed during the measurement When the device has detected the pulse the heart symbol in the display begins to blink Measured result The measured systolic and diastolic blood Pressure values as well as the pulse are now displayed Exa...

Page 15: ...g the patient feels unwell the power button or can be pressed at any time The device then immediately lowers the cuff pressure automatically 6 6 Memory storage and recall of the measurements The blood pressure monitor automatically stores each of the last 120 measurement values By pressing the MEMORY button an average value of the last 3 measurements as well as the last measurement and the further...

Page 16: ...nce of the Heart IHB Irregular Heart Beat Indicator for early Detection This icon indicates that certain pulse irregularities were detected during the measurement In this case the result may deviate from your normal blood pressure repeat the measurement In most cases this is no cause for concern However if the symbol appears on a regular basis e g several times a week with measurements taken daily...

Page 17: ...n then you are advised to visit your doctor If after repeated measurement the AFIB symbol is no longer displayed there is no cause for concern In such case it is recommended to measure again the next day Keep the arm still during measuring to avoid false readings This device may not detect atrial fibrillation in people with pacemakers or defibrillators 9 Error messages malfunctions If an error occ...

Page 18: ...easuring devices see technical data you must consult your specialist dealer or chemist if there are technical problems with the blood pressure instrument Never attempt to repair the instrument yourself Any unauthorized opening of the instrument invalidates all guarantee claims Other possible malfunctions and their elimination If problems occur when using the device the following points should be c...

Page 19: ...vice to extreme temperatures humidity dust or direct sunlight b The cuff contains a sensitive air tight bubble Handle this carefully and avoid all types of straining through twisting or buckling c Clean the device with a soft dry cloth Do not use petrol thinners or similar solvent Spots on the cuff can be removed carefully with a damp cloth and soapsuds The cuff must not be washed d Do not drop th...

Page 20: ... once fitted Do not use the instrument close to strong electromagnetic fields such as mobile telephones or radio installations Do not use the instrument if you think it is damaged or notice anything unusual If the instrument is not going to be used for a prolonged period the batteries should be removed Read the additional safety instructions in the individual sections of this booklet Ensure that c...

Page 21: ...EN1060 1 EN1060 3 EN1060 4 clinical investigation IEC EN 60601 1 11 ANSI AAMI SP10 NIBP IEC80601 2 30 2009 corrigendum 2010 Electrical compatibility Device fulfils the stipulations of the IEC EN 60601 1 IEC EN 60601 1 2 The stipulations of the EU Guidelines 93 42 EEC for Medical Products Class IIa have been fulfilled 13 Remark Some electrical and electrical equipments forbid to abandon and disposa...

Page 22: ...ase V diastolic Display Digital display Measuring range Pressure 30 to 280 mmHg in 1 mmHg increment Pulse 40 to 199 beat minute Static accuracy Pressure 3mmHg Pulse 5 of reading Measuring resolution 1mmHg Inflation Automatic inflation by internal pump Memory function 2 x 120 memories for 2 users SYS DIA Pulse Decompression Constant exhaust valve system Power source 4 size AA alkaline Batteries Rat...

Page 23: ... is intended for use in the electromagnetic environment specified below The customer or the user of this device should assure that it is used in such an environment Electromagnetic Emissions IEC60601 1 2 Emission Test Compliance Electromagnetic Environment RF emission CISPR 11 Group 1 Afib Talk uses RF energy only for internal functions Therefore this RF emission is extremely weak and there is lit...

Page 24: ...ns power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip inUT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip inUT for 25 cycles 5 UT 95 dip inUT for 5 sec Not applicable Mains power quality should be that of a typical commercial or hospital environment I...

Page 25: ... the vicinity of equipment marked with the following symbol Note1 At 80 MHz and 800 MHz the higher frequency range applies Note2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people aField strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile ...

Page 26: ...ing to frequency of transmitter m 150 kHz to 80 MHz d 1 2 p1 2 80 MHz to 800 MHz d 1 2 p1 2 800 MHz to 2 5 GHz d 2 3 p1 2 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be determined using the equation applicable to the frequency of the transmitt...

Reviews: